Why a fixed tolerance fails
MPE grows with length, so a single short gauge skips exactly the long-distance behaviour that structure and environment affect most.
Cover 20%–100% of the range, three runs each, take the worst deviation.
The numbers
Measured vs limit: L=100 mm: 2.1 / 2.83 μm; L=300 mm: 2.6 / 3.50 μm; L=500 mm: 3.2 / 4.17 μm; L=700 mm: 3.9 / 4.83 μm; L=900 mm: 4.6 / 5.50 μm.
All pass; the smallest margin, 0.73 μm, occurs at 900 mm.
Using the margin
Below 20% margin, shorten the re-verification interval and log temperature; two consecutive near-limit results justify compensation or service.
Case workflow, animated
- Confirm the task and datum
- Lay out stations by span
- Acquire readings and convert
- Evaluate and verify
The animation follows the station layout and calculation order of this case. Values are teaching data.
Teaching data. Formal acceptance requires an accredited body under the current ISO 10360-2.