APPLICATION GUIDE

Granite & Precision Bases

Granite plates, straightedges, equipment bases and support surfaces

Methods

Procedure

  1. Define supports
  2. Stabilize temperature
  3. Measure the grid
  4. Check node closure
  5. Evaluate and verify

Support changes can alter form; report support conditions and sampling coverage.

MEASUREMENT SCENARIOS

Typical measurement scenarios in this industry

Each scenario gives the object, tolerance range and method, ready to take to the selector.

Open in the atlas ↗
InspectionFlatness

Grade 000 granite plate inspection

A 1000×1500 mm plate on a 9×7 grid after 24 h soak: rough scan for high points, then fine measurement with a minimum-zone plane.

Metrology autocollimator series↗
ProductionFlatness

Lapping correction closed loop

Measure, map high points, plan the lapping path, measure again. Remove 1–2 μm per pass to avoid destroying the datum.

Metrology autocollimator series↗
InspectionStraightness

Granite straight edge and square

Measure the working face by the pitch method, supporting at the two Bessel points to minimise self-weight sag.

Metrology autocollimator series↗
AssemblyStage Pitch & Yaw

Granite machine bed levelling

Granite has low expansion but is brittle: support evenly at many points, then re-check after 12 h for creep.

Metrology autocollimator series↗
CalibrationMeasurement Uncertainty Budget

Plate certificate and uncertainty

A certificate must state grid, support, temperature, evaluation method and U. A bare 'pass' cannot be reproduced or defended.

Metrology autocollimator series↗
InspectionParallelism

Verification of granite V-blocks and straightedges

Periodic verification of granite V-blocks and straightedges for parallelism and straightness; autocollimator and optical-flat methods cross-check each other.

Metrology autocollimator series↗
InspectionPerpendicularity & Squareness

Granite square perpendicularity

300–1000 mm, typically 3–5 μm; reversal separates instrument and part error.

General-purpose electronic autocollimators↗
InspectionStraightness

Straightedge straightness

1–3 m straightedges; evaluate against end points and correct for sag.

General-purpose electronic autocollimators↗
STEP BY STEP

Understand the measurement step by step

  1. 01

    Define supports

  2. 02

    Stabilize temperature

  3. 03

    Measure the grid

  4. 04

    Check node closure

  5. 05

    Evaluate and verify

Principle schematic · not to scale
1 / 4
  1. Mark supports · Mark support positions, pads and base orientation; record temperature and settling state and establish a reference that does not move with the base.
  2. Record state A · Measure four fixed marks in support state A and retain their coordinates and zero-transfer records.
  3. Reposition to B · Lift and reposition under the approved handling procedure, restore support state B, allow settling and remeasure against the same external reference.
  4. Differences and grid · Compute B−A at each point, inspect support contact near changed locations, and use a complete grid for final surface evaluation.

Teaching animation: follow the stations and reference directions. Do not read dimensions or angles from the schematic.

WORKED EXAMPLE

Reproducing granite-base support after repositioning

Teaching example with assumed data, not a customer project, instrument accuracy claim or acceptance result.

Example task

Teaching demonstration, not a real customer project. All numbers are constructed examples. Check whether marked points change relative to the same external height reference after the base is relocated. The aim is support reproducibility, not a four-point claim about whole-surface flatness.

Setup and preparation

  • Mark support positions, pads and base orientation; record temperature and settling state and establish a reference that does not move with the base.

  • Measure four fixed marks in support state A and retain their coordinates and zero-transfer records.

  • Lift and reposition under the approved handling procedure, restore support state B, allow settling and remeasure against the same external reference.

  • Compute B−A at each point, inspect support contact near changed locations, and use a complete grid for final surface evaluation.

Example data and records

  • State A heights in μm: [0,2,−1,1].

  • State B heights in μm: [0,1,−2,1].

  • Point differences B−A in μm: [0,−1,−1,0].

Calculation steps

Δzᵢ=zBᵢ−zAᵢ; max|Δzᵢ|=1 μm and the difference range is 0−(−1)=1 μm.

Interpretation

The maximum absolute change at the four sampled positions is 1 μm. Its significance depends on reference stability and uncertainty; it does not establish 1 μm flatness of the entire base.

Verification and pitfalls

  • Independently verify the external reference: its motion could mimic base motion.

  • Keep support points, pads and fastening consistent, recording unavoidable changes.

  • Complete thermal settling and full-grid verification before judging the final installation state.