APPLICATION GUIDE

Metrology & R&D

Angle references, surface plates, reference platforms and research instrument prototypes.

Methods

Procedure

  1. Define the measurand
  2. Establish traceability
  3. Measure and repeat
  4. Evaluate uncertainty
  5. Review the report

Use validated procedures and calibrated references for calibration or conformity claims.

MEASUREMENT SCENARIOS

Typical measurement scenarios in this industry

Each scenario gives the object, tolerance range and method, ready to take to the selector.

Open in the atlas ↗
CalibrationSmall Angles & Indexing Error

Angle gauge and polygon verification

Use the autocollimator as the readout with a rotary stage, comparing face by face at 20±0.5 °C, three repeats per face.

Goniometers↗
CalibrationFlatness

Cast-iron and granite plate flatness

Grid, diagonal or circular layouts. Grade 0 plates are often 3–8 μm; soak at least 12 h and keep away from sunlight and air outlets.

Metrology autocollimator series↗
CalibrationPerpendicularity & Squareness

Square and box squareness

Reversal removes instrument bias: average the two readings of a forward and reversed setup so the autocollimator residual drops out.

Metrology autocollimator series↗
CalibrationStraightness

CMM guideway and column geometry

Assess straightness per axis, squareness between axes and angular motion; acceptance follows the ISO 10360 or GB/T 16857 item set.

Metrology autocollimator series↗
CalibrationMeasurement Uncertainty Budget

Measurement uncertainty budget and report

List indication error, span, mirror angle, temperature and repeatability, combine by root-sum-square, and state method and datum in the report.

Metrology autocollimator series↗
CalibrationRotary Positioning & Reversal

Indexing check of roundness tester spindles

Verify angular sampling intervals of roundness tester spindles with the polygon method so harmonic analysis rests on a correct angle base; annual check.

Goniometers↗
CalibrationSmall Angles & Indexing Error

Indication error comparison of autocollimators

Metrology labs check in-service autocollimators with angle gauges or small-angle generators; interim checks typically every 6–12 months.

Metrology autocollimator series↗
CalibrationSmall Angles & Indexing Error

Angle transfer

Transfer from polygon or angle blocks to working instruments, logging uncertainty at each step.

Goniometers↗
CalibrationFlatness

Surface plate calibration and lapping allowance

Grid measurement gives a lapping map instead of uniform removal.

General-purpose electronic autocollimators↗

Recommended instruments

STEP BY STEP

Understand the measurement step by step

  1. 01

    Define the measurand

  2. 02

    Establish traceability

  3. 03

    Measure and repeat

  4. 04

    Evaluate uncertainty

  5. 05

    Review the report

u_c = √Σuᵢ²U = k · u_cPrinciple schematic · not to scale
1 / 4
  1. Define model · Define the measurand, correction signs and measurement model. For this example all three sensitivity coefficients are +1.
  2. Standardize components · Convert certificate, repeated-reading and environmental information into standard uncertainties. Do not directly use expanded uncertainties or limit values as standard components.
  3. Combine and expand · Check whether components are correlated, then combine independent components by root sum of squares.
  4. Record budget limits · Explain the coverage-factor choice and record the result, unit, factor, applicable range and excluded influences together.

Teaching animation: follow the stations and reference directions. Do not read dimensions or angles from the schematic.

WORKED EXAMPLE

Metrology laboratory: a simplified angle-comparison uncertainty budget

Teaching example with assumed data, not a customer project, instrument accuracy claim or acceptance result.

Example task

Teaching demonstration, not a real customer project. Three independent components, already expressed as standard uncertainties, are combined for a defined small-angle comparison. The numbers are teaching inputs, not the calibration capability of any instrument or laboratory.

Setup and preparation

  • Define the measurand, correction signs and measurement model. For this example all three sensitivity coefficients are +1.

  • Convert certificate, repeated-reading and environmental information into standard uncertainties. Do not directly use expanded uncertainties or limit values as standard components.

  • Check whether components are correlated, then combine independent components by root sum of squares.

  • Explain the coverage-factor choice and record the result, unit, factor, applicable range and excluded influences together.

Example data and records

  • Reference-calibration component u1=0.08″.

  • Standard repeatability component for the result u2=0.06″; environmental component u3=0.03″.

  • Teaching assumptions: independent components and unit sensitivity coefficients; k=2 is used for illustration.

Calculation steps

uc=sqrt(0.08²+0.06²+0.03²)=sqrt(0.0109)=0.104403″; U=kuc=2×0.104403=0.208806″.

Interpretation

Under the stated assumptions, combined standard uncertainty is approximately 0.104403″ and illustrative expanded uncertainty is 0.208806″. A factor k=2 does not automatically prove an exact 95% coverage probability.

Verification and pitfalls

  • Include covariance terms when inputs are correlated instead of retaining the independent formula.

  • Check whether repeatability applies to one reading or a mean, avoiding duplication or omission.

  • A real certificate needs traceability, a complete model and a justified coverage factor; do not copy these example values.