Knowledge & technical search
Principles, error sources and practical guidance in one place.
Local technical search across the curated site content.
Understand these essentials
Why is resolution not accuracy?
Resolution describes the smallest displayed or distinguishable change. Accuracy also depends on calibration, range, environment and the measurement system.
Least squares or minimum zone?
Least squares minimizes squared residuals. Minimum zone minimizes the separation of enclosing boundaries. Their results are not interchangeable.
Can one line establish flatness?
No. Surface evaluation needs a sampling network and a common height reference, with limitations stated for unmeasured regions.
How can drift be controlled?
Stabilize supports and environment, control vibration and thermal disturbances, record temperature, and use repeated reference readings and reverse runs.
Understand the principle, then the case
18 notes explain where the angle comes from and how to read the specifications. The case library is organised by task and industry.
How an autocollimator turns angle into a reading
An autocollimator illuminates a plane mirror with a parallel beam. The returning beam is focused again by the objective, and the image position shifts with the mirror angle. Every autocollimation measurement rests on this relationship.
PRINCIPLEDual-axis measurement, field of view and targets
A dual-axis autocollimator reads two angular directions in one setup and reduces repeated adjustment. Field of view, resolution and working distance, however, trade off against each other.
PRINCIPLEReading resolution, indication error and repeatability
Resolution is usually the first number on a datasheet, yet usability depends on indication error within a stated range and on repeatability. The three are not interchangeable.