Principle

Uncertainty belongs to a defined measurand and model; distinguish applied corrections from remaining uncertainty.

Preparation

Define the measurand, units, input sources, corrections, correlations and coverage requirements.

Workflow diagram

01Write the relationship between the measurand and input02List estimates03Calculate sensitivity coefficients and include relevan04Combine standard uncertainty and justify the coverage 05Report result

Procedure

  1. Write the relationship between the measurand and inputs.
  2. List estimates, distributions, standard uncertainties and evidence.
  3. Calculate sensitivity coefficients and include relevant correlations.
  4. Combine standard uncertainty and justify the coverage factor.
  5. Report result, expanded uncertainty, units, coverage factor and conditions.

Calculation

For independent inputs under first-order propagation: uc²=Σ(cᵢuᵢ)². Add covariance terms for correlated inputs. Expanded uncertainty U=kuc requires a justified k.

Illustrative example

Schematic · not measured data
Principle and workflow schematic. Numbers are worked examples, not customer or instrument test results.

Illustration: independent standard uncertainties of 0.10″ and 0.05″ give uc≈0.112″. If k=2 is justified, U≈0.224″.

Limitations

  • Manufacturer resolution is not the total uncertainty.
  • Root-sum-of-squares is insufficient when relevant correlations exist.

Reporting

Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.

Method reference ↗ Method reference ↗ Method reference ↗ 
STEP BY STEP

Understand the measurement step by step

u_c = √Σuᵢ²U = k · u_cPrinciple schematic · not to scale
1 / 4
  1. Define model · Write the measurement model for angular deviation; assume sensitivity coefficients have already been applied and the three contributions are uncorrelated.
  2. Enter standard contributions · Enter standard contributions from repeats, the reference and environment/mounting; convert limits or expanded values appropriately before use.
  3. Combine variances · Sum the squared standard contributions and take the square root, showing each variance contribution in the animation.
  4. State coverage factor · Calculate U with k=2 only when that factor is justified; report the factor, method and assumptions without asserting exactly 95% coverage for every distribution.

Teaching animation: follow the stations and reference directions. Do not read dimensions or angles from the schematic.

WORKED EXAMPLE

A three-component angular uncertainty budget

Teaching example with assumed data, not a customer project, instrument accuracy claim or acceptance result.

Example task

Teaching demonstration, not a real customer project: combine three standard-uncertainty contributions already expressed in output-angle units. Resolution, accuracy and expanded uncertainty are distinct concepts.

Setup and preparation

  • Write the measurement model for angular deviation; assume sensitivity coefficients have already been applied and the three contributions are uncorrelated.

  • Enter standard contributions from repeats, the reference and environment/mounting; convert limits or expanded values appropriately before use.

  • Sum the squared standard contributions and take the square root, showing each variance contribution in the animation.

  • Calculate U with k=2 only when that factor is justified; report the factor, method and assumptions without asserting exactly 95% coverage for every distribution.

Example data and records

  • Standard contributions: repeated readings 0.08 arcsec, reference 0.05 arcsec, environment/mounting 0.03 arcsec.

  • Sensitivity coefficients are already included; correlations are assumed zero.

  • Demonstration coverage factor k=2; actual use requires a basis for its selection.

Calculation steps

uc=√(0.08²+0.05²+0.03²)=√0.0098=0.098995 arcsec; U=2uc=0.197990 arcsec. Variance shares are approximately 65.31%, 25.51% and 9.18%.

Interpretation

Under these assumptions, combined standard uncertainty is about 0.0990 arcsec and expanded uncertainty for k=2 is about 0.1980 arcsec. These are worked-budget values, not a model-specific calibration or accuracy promise.

Verification and pitfalls

  • Include covariance terms when contributors are correlated; a shared reference is not automatically independent.

  • Document distributions, divisors, sensitivity coefficients and degrees of freedom.

  • Check first-order propagation and coverage assumptions for nonlinear or non-normal cases.