Home / Rotary table and indexing head calibration
DOWNLOADABLE MEASUREMENT PLAN

Rotary table and indexing head calibration

Full-circle indexing error and repeatability of NC tables, indexing heads and multi-tooth tables for acceptance, compensation and repair verification.

Instruments

Procedure

Calculation

Error = table reading − nominal polygon angle − zero offset. Closure checks consistency.

Reporting

Limitations

Polygon grade sets the floor; radial and axial runout are separate.

STEP BY STEP

Understand the measurement step by step

  1. 01

    Confirm polygon faces, eccentricity and start angle; define stations.

  2. 02

    Mount the polygon coaxially and align the autocollimator over the full circle.

  3. 03

    Index face by face and record table and autocollimator readings.

  4. 04

    Compute indexing error and closure; run both directions to separate backlash.

  5. 05

    Build a compensation table if required and re-measure.

Principle schematic · not to scale
1 / 5
  1. Step 1 · Confirm polygon faces, eccentricity and start angle; define stations.
  2. Step 2 · Mount the polygon coaxially and align the autocollimator over the full circle.
  3. Step 3 · Index face by face and record table and autocollimator readings.
  4. Step 4 · Compute indexing error and closure; run both directions to separate backlash.
  5. Step 5 · Build a compensation table if required and re-measure.

Teaching animation: follow the stations and reference directions. Do not read dimensions or angles from the schematic.

WORKED EXAMPLE

Worked 12-face polygon index error with closure check

Teaching example with assumed data, not a customer project, instrument accuracy claim or acceptance result.

Example task

Teaching example: the table is compared with a 12-face polygon every 30°; the sum of all deviations should close to zero and is used as a consistency check.

Setup and preparation

  • Confirm polygon faces, eccentricity and start angle; define stations.

  • Mount the polygon coaxially and align the autocollimator over the full circle.

  • Index face by face and record table and autocollimator readings.

  • Compute indexing error and closure; run both directions to separate backlash.

Example data and records

  • Index deviations in arc-seconds: +0.0, +0.4, -0.3, +0.2, -0.1, +0.5, -0.4, +0.1, -0.2, +0.3, -0.5, +0.0.

  • Closure 0″, peak-to-peak 1″.

  • Polygon grade and autocollimator resolution are logged separately; together they set the floor.

Calculation steps

Index error = table reading − nominal polygon angle − start zero; the twelve terms close to 0″. A non-zero closure means a zero drift or a reading error.

Interpretation

Peak-to-peak index error is 1″ with a near-sinusoidal pattern, typical of eccentric mounting.

Verification and pitfalls

  • Polygon and table must be coaxial; eccentricity appears as the first harmonic.

  • Read each position in both directions to separate backlash from true index error.

  • Temperature changes expand polygon and table differently; log it during long runs.