APPLICATION GUIDE

Education

Optical benches, metrology teaching platforms and assembly training equipment.

Methods

Procedure

  1. Explain the principle
  2. Assemble the optical path
  3. Record readings
  4. Compare evaluation methods
  5. Review errors

Teaching calculations and sample plots are not calibration certificates.

MEASUREMENT SCENARIOS

Typical measurement scenarios in this industry

Each scenario gives the object, tolerance range and method, ready to take to the selector.

Open in the atlas ↗
TeachingSmall Angles & Indexing Error

Autocollimation demonstration

A beamsplitter and flat mirror show parallel out and focus-plane return; students turn the drum and read arcseconds to build intuition.

General-purpose electronic autocollimators↗
TeachingStraightness

Pitch-method straightness exercise

Eleven points at 100 mm on a 1 m rail: hand-accumulate, fit least squares, then verify in software to see angles integrate into a profile.

General-purpose electronic autocollimators↗
TeachingFlatness

Grid-method flatness exercise

A 5×5 grid on a 300×300 mm plate: groups measure, swap data and evaluate separately to compare methods.

General-purpose electronic autocollimators↗
TeachingMeasurement Uncertainty Budget

Error sources and uncertainty project

Students design a measurement, list at least six error sources, combine them, and explain why their result differs from a classmate's.

General-purpose electronic autocollimators↗
TeachingOptical Alignment

Optical assembly and alignment practicum

From seating a lens to aligning the group, accept at element, assembly and system level so students locate error before adjusting.

Wide-field electronic autocollimators↗
TeachingRepeatability & Re-setup Verification

Repeatability and reproducibility study

Three operators, ten repeats each: compare within and between groups to see that a good instrument is not a good system.

General-purpose electronic autocollimators↗
TeachingMeasurement Uncertainty Budget

Teaching lab on uncertainty evaluation

Using prism dihedral measurement as the vehicle, train repeatability statistics, component combination and k=2 expanded uncertainty reporting in 2 teaching hours.

General-purpose electronic autocollimators↗

Recommended instruments

STEP BY STEP

Understand the measurement step by step

  1. 01

    Explain the principle

  2. 02

    Assemble the optical path

  3. 03

    Record readings

  4. 04

    Compare evaluation methods

  5. 05

    Review errors

x̄12345Repeated readings scatter about the mean; not a drift curveReading numberPrinciple schematic · not to scale
1 / 4
  1. Fix conditions · Fix the optical path and state the angular units. Keep reflector, operator and environment unchanged within this group to avoid mixing in reinstallation effects.
  2. Sequential readings · Collect five readings after a predefined stabilization procedure and retain their sequence; do not delete a reading merely because it looks large.
  3. Calculate statistics · Calculate the mean and the sample standard deviation with denominator n−1, then calculate the Type A uncertainty of the mean under an independent, identically distributed assumption.
  4. Compare groups · Conduct a separate reinstallation exercise and compare group means and within-group dispersion. Do not indiscriminately combine the two experiments as repeatability.

Teaching animation: follow the stations and reference directions. Do not read dimensions or angles from the schematic.

WORKED EXAMPLE

Teaching lab: resolution, repeatability and uncertainty of the mean

Teaching example with assumed data, not a customer project, instrument accuracy claim or acceptance result.

Example task

Teaching demonstration, not a real customer project. Students repeatedly read one small angle with the reflector fixed and without reinstallation, distinguishing sample standard deviation from the Type A standard uncertainty of a mean. Display digits are not treated as accuracy.

Setup and preparation

  • Fix the optical path and state the angular units. Keep reflector, operator and environment unchanged within this group to avoid mixing in reinstallation effects.

  • Collect five readings after a predefined stabilization procedure and retain their sequence; do not delete a reading merely because it looks large.

  • Calculate the mean and the sample standard deviation with denominator n−1, then calculate the Type A uncertainty of the mean under an independent, identically distributed assumption.

  • Conduct a separate reinstallation exercise and compare group means and within-group dispersion. Do not indiscriminately combine the two experiments as repeatability.

Example data and records

  • Five synthetic fixed-condition readings: 0.20″, 0.30″, 0.10″, 0.20″, 0.20″.

  • Sample size n=5; calibration corrections and environmental uncertainties are not supplied.

  • Independence and identical distribution are assumed only for this statistics lesson.

Calculation steps

Mean x̄=0.20″; sample standard deviation s=sqrt[0.02/(5−1)]=0.070711″; Type A standard uncertainty of the mean uA=s/√5=0.031623″.

Interpretation

The group spread is described by s≈0.070711″ and the statistical component for its mean by uA≈0.031623″. Neither value replaces overall accuracy or a complete measurement-uncertainty budget.

Verification and pitfalls

  • Inspect the sequence for drift; correlated readings cannot automatically use s/√n.

  • Resolution describes display or quantization, not calibrated accuracy.

  • Document any outlier rule and retain all original readings when no technical exclusion is justified.