MICROCRE / OPTICAL METROLOGY400-812-7833  ·  SHANGHAI
研润 WHW
MICROCREPRECISION IN EVERY ANGLE
Home / Laboratory Platforms and Foundation Equipment
APPLICATION GUIDE

Laboratory Platforms and Foundation Equipment

Optical tables, vibration-isolated bases, precision instrument mounts, rail stages, reference artefacts in temperature-controlled laboratories

Methods

Procedure

  1. Stabilise the environment
  2. Level the platform
  3. Measure flatness
  4. Check the stage
  5. Record and re-verify

Geometric checks do not replace vibration-transmissibility tests, temperature and humidity stability verification, or the instrument's own metrological calibration.

STEP BY STEP

Understand the measurement step by step

  1. 01

    Stabilise the environment

  2. 02

    Level the platform

  3. 03

    Measure flatness

  4. 04

    Check the stage

  5. 05

    Record and re-verify

Principle schematic · not to scale
1 / 4
  1. Mark nodes · Mark five nodes along a line on the table top surface at 50 mm spacing (total 200 mm) and define a positive angle as an upward step to the next node; the reflector-base span equals the 50 mm sampling step.
  2. Align optics · Set up the autocollimator using the equal-height line through the table supports for coarse alignment, keep the readings near mid-range and confirm that the reflector base keeps its orientation while it is moved from node to node.
  3. Integrate · Integrate the intervals into five node heights, use the line through the first and last nodes as the reference, compute the residuals and then repeat the run in reverse over the same nodes.
  4. Reverse run · Repeat the run in reverse over the same nodes, compare the two residual profiles and keep the raw angles together with before-and-after adjustment records.

Teaching animation: follow the stations and reference directions. Do not read dimensions or angles from the schematic.

WORKED EXAMPLE

Teaching example: flatness of an optical table

Teaching example with assumed data, not a customer project, instrument accuracy claim or acceptance result.

Example task

Teaching demonstration, not a real customer project: a 200 mm section of an optical table shows how interval angles from autocollimation are reconstructed into a height profile. The values are teaching assumptions and are not acceptance limits for laboratory platforms.

Setup and preparation

  • Mark five nodes along a line on the table top surface at 50 mm spacing (total 200 mm) and define a positive angle as an upward step to the next node; the reflector-base span equals the 50 mm sampling step.

  • Set up the autocollimator using the equal-height line through the table supports for coarse alignment, keep the readings near mid-range and confirm that the reflector base keeps its orientation while it is moved from node to node.

  • Read and retain the raw angle of every interval rather than a smoothed display value, and log ambient temperature, time of measurement, and conditions such as building vibration and air-conditioning draught.

  • Integrate the intervals into five node heights, use the line through the first and last nodes as the reference, compute the residuals and then repeat the run in reverse over the same nodes.

Example data and records

  • Interval angles: 0.5″, 0.3″, −0.4″, 0.2″.

  • Span and sampling step: 50 mm, four intervals; the first node is defined as 0 μm.

  • Cumulative node heights: 0.000, 0.121, 0.194, 0.097, 0.145 μm; residuals against the endpoint line: 0.000, 0.085, 0.121, −0.012, 0.000 μm.

Calculation steps

For each interval Δh = 50×1000×tan(θ·π/648000) μm. With the four angles this gives Δh ≈ 0.121, 0.073, −0.097, 0.048 μm; cumulative heights are 0.000, 0.121, 0.194, 0.097, 0.145 μm, residuals after removing the endpoint line are 0.000, 0.085, 0.121, −0.012, 0.000 μm and the peak-to-valley value is 0.133 μm.

Interpretation

For the given sampling and endpoint reference the residual peak-to-valley of a line on the table top surface is about 0.133 μm. It describes only the static geometry of this 200 mm line and is not an acceptance conclusion for laboratory platforms.

Verification and pitfalls

  • Confirm that the reflector span matches the actual sampling step; the simple integration model cannot be applied when they differ.

  • A reverse run shows repeatability only and does not remove contact deformation or thermal drift of an optical table; building vibration and air-conditioning draught must be assessed separately.

  • Any report must state the reference, evaluation algorithm, tolerance and measurement uncertainty.