Principle
An electronic level reads the base tilt relative to local vertical, an autocollimator the reflector tilt relative to its optical axis; both output angles and share the same integration formula for profile reconstruction. The level is immune to beam drift but sensitive to support condition and vibration; the autocollimator is gravity-insensitive but depends on optical stability. Comparing two profiles measured on the same line: agreement indicates controlled systematic errors, while systematic differences point to instrument datum or environmental causes.
Preparation
Verify the calibration status and resolution of both instruments (typically 0.01″–1″ class). Plan one common line with identical stations and span, ideally using one shared base or geometrically matched bases. Check level drift and autocollimator return light, and take readings simultaneously or alternately to suppress environmental drift.
Workflow diagram
Procedure
- Define the line, stations and span; mark positions and record temperature.
- Complete a full traverse with one instrument, e.g. the autocollimator, logging every station angle.
- Repeat on the same stations with the electronic level, keeping identical positions.
- Integrate both angle sequences with the same formula and span to obtain two height profiles.
- Overlay the profiles; compute point-wise differences, their standard deviation and any trend.
- If differences exceed the agreed limit, investigate instrument datum, support, temperature and vibration, then issue the comparison conclusion after re-checking.
Calculation
Both instruments integrate with Δh=L·tan(θ″·π/648000). The profile difference is δi=zA,i−zB,i; consistency is judged by the standard deviation s=√[Σδi²/(n−1)] with any trend reported separately. A linear component in δ suggests datum drift of one instrument; a random component suggests stochastic effects. The comparison limit combines both uncertainties: Uc≈√(UA²+UB²).
Illustrative example
Teaching data, not measured: span 100 mm, eight stations. The autocollimator profile is [0, +0.5, +0.9, +1.2, +1.0, +0.7, +0.3, 0] μm and the level profile [0, +0.4, +1.0, +1.3, +1.1, +0.8, +0.4, 0] μm, with point differences of about 0.1 μm and no systematic trend. The example illustrates a good-agreement verdict.
Limitations
- Non-coincident station positions mix layout error into the differences.
- Levels are sensitive to thermal gradients and draughts; cross-check under the measurement environment.
- A single comparison does not replace periodic calibration; it verifies consistency, not indicated values.
Reporting
Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.