Principle
Uncertainty belongs to a defined measurand and model; distinguish applied corrections from remaining uncertainty.
Preparation
Define the measurand, units, input sources, corrections, correlations and coverage requirements.
Workflow diagram
Procedure
- Write the relationship between the measurand and inputs.
- List estimates, distributions, standard uncertainties and evidence.
- Calculate sensitivity coefficients and include relevant correlations.
- Combine standard uncertainty and justify the coverage factor.
- Report result, expanded uncertainty, units, coverage factor and conditions.
Calculation
For independent inputs under first-order propagation: uc²=Σ(cᵢuᵢ)². Add covariance terms for correlated inputs. Expanded uncertainty U=kuc requires a justified k.
Illustrative example
Illustration: independent standard uncertainties of 0.10″ and 0.05″ give uc≈0.112″. If k=2 is justified, U≈0.224″.
Limitations
- Manufacturer resolution is not the total uncertainty.
- Root-sum-of-squares is insufficient when relevant correlations exist.
Reporting
Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.