CMM MPE acceptance: judging five lengths

The ISO 10360-2 MPE limit varies with length. Five measured lengths compared with the formula show whether the machine is still in acceptance state.

ObjectBridge CMM, 900 mm range
MethodISO 10360-2, five gauges, three runs each
MPEMPE_E0 = 2.5 + L/300 μm
ResultWorst 4.6 μm at 900 mm against 5.5 μm
h_i = sum L tan(theta)

Why a fixed tolerance fails

MPE grows with length, so a single short gauge skips exactly the long-distance behaviour that structure and environment affect most.

Cover 20%–100% of the range, three runs each, take the worst deviation.

The numbers

Measured vs limit: L=100 mm: 2.1 / 2.83 μm; L=300 mm: 2.6 / 3.50 μm; L=500 mm: 3.2 / 4.17 μm; L=700 mm: 3.9 / 4.83 μm; L=900 mm: 4.6 / 5.50 μm.

All pass; the smallest margin, 0.73 μm, occurs at 900 mm.

Using the margin

Below 20% margin, shorten the re-verification interval and log temperature; two consecutive near-limit results justify compensation or service.

Case workflow, animated

SourceClear aperture → max distanceLonger distance: larger spot, weaker returnSchematic · not to scale · not measured data
1 / 4
  1. Confirm the task and datum
  2. Lay out stations by span
  3. Acquire readings and convert
  4. Evaluate and verify

The animation follows the station layout and calculation order of this case. Values are teaching data.

ConclusionAll five lengths pass, with the smallest margin 0.73 μm at the longest length — the trend to watch.

Teaching data. Formal acceptance requires an accredited body under the current ISO 10360-2.