Method: locate by two reflections
Place a reflective target at the path entrance and at the nozzle seat, and aim the autocollimator at each in turn. The reading difference shows the path bending error; adjust the mirror mounts until both targets read on the same datum.
This is faster and quantitative compared with burn-spot methods: every adjustment shows an angle change instead of guessing from a burn mark.
Procedure: coarse, fine, re-check
Coarse-align with the supplied alignment film until the return image is visible on the target, then fine-tune with the autocollimator. Over a 2.5 m path, 1″ is about 12 μm at the nozzle; the 0.05 mm target leaves about 4″ of angular margin.
Teaching example: initial two-target difference 11.3″, converging to 2.1″ after two adjustment rounds, about 0.025 mm coaxial deviation.
Verification: close the loop with cut quality
After the angular alignment passes, cut a standard coupon to verify: kerf perpendicularity and dross are the final evidence of coaxiality. Instrument alignment and process verification are both required.
Record both reading sets before/after alignment together with coupon photos as annexes to the maintenance report.
Case workflow, animated
- Confirm the task and datum
- Lay out stations by span
- Acquire readings and convert
- Evaluate and verify
The animation follows the station layout and calculation order of this case. Values are teaching data.
Case data are teaching assumptions. Laser equipment alignment must be performed by authorised personnel under power-off and protective conditions; coaxiality acceptance follows the manufacturer service manual.
Recommended instruments

Dual-axis autocollimator, accuracy 0.1″, range 30 m

Dual-axis autocollimator, accuracy 1″, range 40 m