Principle

Geometric measurement supplies feedback; contact, stiffness and locking affect adjustment response.

Preparation

Define permitted adjustment points, assembly constraints, supports and verification conditions; preserve baseline data.

Workflow diagram

01Measure the full travel and check repeatability before02Separate orientation03Make controlled small adjustments under the assembly p04Remeasure the same lines and conditions without changi05Lock the assembly

Procedure

  1. Measure the full travel and check repeatability before adjustment.
  2. Separate orientation, form and local contact anomalies.
  3. Make controlled small adjustments under the assembly procedure and record each action.
  4. Remeasure the same lines and conditions without changing the evaluation reference.
  5. Lock the assembly, restore the specified load and verify again.

Calculation

With identical stations and reference, Δrᵢ=rᵢ,after−rᵢ,before. Keep global direction and form changes separate.

Illustrative example

Schematic · not measured data
Principle and workflow schematic. Numbers are worked examples, not customer or instrument test results.

Illustration: a point changing from +4 to +1 μm in the same reference changes by −3 μm; other stations and the locked state still require checks.

Limitations

  • Do not machine or scrape based on an unverified single trace.
  • Changing one support can alter other positions and load distribution.

Reporting

Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.

Method reference ↗ Method reference ↗ Method reference ↗ 
STEP BY STEP

Understand the measurement step by step

Principle schematic · not to scale
1 / 4
  1. Save initial condition · Define five common stations, support conditions and an endpoint-line reference; retain the initial profile and fastening record.
  2. Adjust under procedure · Follow an approved mechanical adjustment procedure; the animation identifies adjustment locations without inventing screw torques or adjustment amounts.
  3. Repeat same reference · Recheck the provisional profile against the same reference and calculate station-by-station changes, without selecting a more favourable reference.
  4. Verify after locking · Lock according to the approved procedure and remeasure for rebound; supplement with orientation, load and temperature checks under intended operating conditions.

Teaching animation: follow the stations and reference directions. Do not read dimensions or angles from the schematic.

WORKED EXAMPLE

Before adjustment, after adjustment and after locking

Teaching example with assumed data, not a customer project, instrument accuracy claim or acceptance result.

Example task

Teaching demonstration, not a real customer project: compare three guideway profiles using the same stations and endpoint reference, showing why an initially improved profile is not necessarily the final stable state.

Setup and preparation

  • Define five common stations, support conditions and an endpoint-line reference; retain the initial profile and fastening record.

  • Follow an approved mechanical adjustment procedure; the animation identifies adjustment locations without inventing screw torques or adjustment amounts.

  • Recheck the provisional profile against the same reference and calculate station-by-station changes, without selecting a more favourable reference.

  • Lock according to the approved procedure and remeasure for rebound; supplement with orientation, load and temperature checks under intended operating conditions.

Example data and records

  • Endpoint-referenced residuals before adjustment: [0,6,9,5,0] μm.

  • After provisional adjustment: [0,2,3,2,0] μm.

  • After locking: [0,2.5,3.5,2.5,0] μm.

Calculation steps

Provisional change r_after−r_before=[0,−4,−6,−3,0] μm. The three peak-to-valley ranges are 9, 3 and 3.5 μm; locking raises the middle point by 0.5 μm.

Interpretation

The worked provisional range improves from 9 to 3 μm but becomes 3.5 μm after locking. Report the specified final working condition, not only the best run.

Verification and pitfalls

  • Keep stations, line and evaluation reference identical.

  • Follow equipment maintenance requirements and preserve stiffness and contact conditions.

  • No tolerance or uncertainty is supplied, and no actual repair improvement is promised.