Principle
A level measures angle to gravity; an autocollimator measures angle to the optical axis. They are different datums.
Preparation
Confirm feet, shims and grouting; allow the specified settling and preload. Record temperature and floor vibration.
Workflow diagram
Procedure
- Load the feet as specified and check shim contact and preload.
- Coarse-level in both directions with an electronic level.
- Re-check at several positions to separate overall tilt from local twist.
- Establish the optical datum with an autocollimator and record zero and sign convention.
- File the levelling record, foot load state and re-check date.
Calculation
1 mm/m ≈ 206.3″; 1″ ≈ 4.85 μm/m. Height difference = reading difference × span.
Illustrative example
Example: 0.02 mm/m and 0.01 mm/m correspond to about 4.1″ and 2.1″, or 40 μm and 20 μm over 2 m.
Limitations
- Treating overall tilt as straightness error.
- Levelling one direction only and missing twist.
- Measuring before the structure has settled.
Reporting
Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.