Home / Thermal Drift & Stability
MEASUREMENT METHOD

Thermal Drift & Stability

Separate time drift from positional error and record thermal effects on angular readings.

Principle

With position held fixed, time-varying readings characterize combined system and environmental stability.

Preparation

Use a fixed reflector, timestamps and temperature sensors; record power state, heat sources and airflow without moving the setup.

Workflow diagram

01Define the observation period02Record angle03Mark power04Inspect trends05Repeat observations to establish usable stability cond

Procedure

  1. Define the observation period, sampling and operating state.
  2. Record angle, temperature and time with the setup fixed.
  3. Mark power, operator and ventilation events.
  4. Inspect trends, short-term noise and correlations without assuming causation.
  5. Repeat observations to establish usable stability conditions and document any compensation or waiting policy.

Calculation

Relative drift d(t)=θ(t)−θ(t₀). A fitted time slope may be reported in arcsec/hour together with duration and operating conditions.

Illustrative example

Schematic · not measured data
Principle and workflow schematic. Numbers are worked examples, not customer or instrument test results.

Illustration: a change from 0.1″ to 0.4″ in one hour gives a 0.3″ endpoint change; two points do not prove linear drift.

Limitations

  • Do not prescribe one universal warm-up time without evidence.
  • Short-term averaging does not automatically remove slow drift.

Reporting

Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.

Method reference ↗ Method reference ↗ Method reference ↗ 
STEP BY STEP

Understand the measurement step by step

Principle schematic · not to scale
1 / 4
  1. Fix reference · Keep instrument, reflector and supports fixed; schedule one observation every ten minutes from 0 to 40 minutes and prepare an event log.
  2. Record by time · Enter the illustrative angles and reserve fields in an actual plan for temperature, power-on, contact and airflow events.
  3. Calculate drift and trend · Subtract the initial reading; calculate both the endpoint change and the least-squares slope through all five observations.
  4. Verify longer windows · Animate a time cursor and data points while showing the limited observation window; plan longer and repeat power-on observations separately.

Teaching animation: follow the stations and reference directions. Do not read dimensions or angles from the schematic.

WORKED EXAMPLE

A 40-minute fixed-reference drift observation

Teaching example with assumed data, not a customer project, instrument accuracy claim or acceptance result.

Example task

Teaching demonstration, not a real customer project: simulated readings over 40 minutes distinguish net drift from a fitted trend. Without a causal temperature model, the change cannot all be attributed to temperature.

Setup and preparation

  • Keep instrument, reflector and supports fixed; schedule one observation every ten minutes from 0 to 40 minutes and prepare an event log.

  • Enter the illustrative angles and reserve fields in an actual plan for temperature, power-on, contact and airflow events.

  • Subtract the initial reading; calculate both the endpoint change and the least-squares slope through all five observations.

  • Animate a time cursor and data points while showing the limited observation window; plan longer and repeat power-on observations separately.

Example data and records

  • Time: [0,10,20,30,40] min.

  • Angle: [0.10,0.20,0.35,0.45,0.50] arcsec.

  • Relative drift: [0,0.10,0.25,0.35,0.40] arcsec.

Calculation steps

Endpoint change=0.50−0.10=0.40 arcsec; endpoint-average rate=0.40/40=0.0100 arcsec/min. The five-point OLS slope Σ(t−mean t)(θ−mean θ)/Σ(t−mean t)² is 0.0105 arcsec/min.

Interpretation

The illustrative window has a net change of 0.40 arcsec and fitted slope of 0.0105 arcsec/min. A short linear fit does not establish all-day drift, warm-up time or a temperature coefficient.

Verification and pitfalls

  • Thermal motion of the fixed reference can itself affect the readings.

  • Preserve timestamps and events to identify contact, airflow and power interruptions.

  • Correlation with temperature is not causation; determine a coefficient through a separate designed experiment.