Home / Minimum-Zone Evaluation of Form Errors
MEASUREMENT METHOD

Minimum-Zone Evaluation of Form Errors

Derive the minimum-condition error value of a reconstructed line or surface profile for acceptance and arbitration. When a drawing specifies form tolerances such as straightness or flatness, the minimum-zone value is the acceptance criterion. It serves as the final evaluation when endpoint or least-squares references are disputed.

Principle

The evaluation reference is an ideal geometric element constructed from the data, not the physical measurement datum. The minimum-zone method searches two parallel lines or planes that contain all points with minimum separation. For a straightness profile the criterion is alternating contacts of high and low points on the reference; for a plane it is a crossing pattern of three high (low) and one low (high) points. Numerically it is a linear-programming or rotation-search problem: rotate and translate the reference to minimise the peak-to-valley distance.

Preparation

Prepare the integrated, unified point data and check span, numbering and sign conventions. Select a constrained-optimisation tool or verified evaluation program and validate it on simple cases first. Keep a copy of the raw data so the evaluation is reproducible.

Workflow diagram

01Prepare data02Initial fit03Rotate and search04Check criterion05Report error06Compare and conclude

Procedure

  1. Import the unified coordinates and heights; verify completeness and consistent units.
  2. Start from a least-squares line or plane and compute residuals and the peak-to-valley distance.
  3. Adjust rotation and translation parameters iteratively to narrow the containment band.
  4. At the minimum band, verify that the contact points satisfy the alternating (crossing) criterion.
  5. Record the optimum reference parameters, contact points and the minimum-zone value.
  6. Compare with endpoint and least-squares results and report the conclusion with reproducibility information.

Calculation

With reference z=a·x+b (plane: z=a·x+b·y+c), residuals are ri=zi−(a·xi+b). The minimum-zone problem is min over a,b of [max(ri)−min(ri)], solvable as linear programming. In manual rotation, Δa is chosen so that the current highest point and the runner-up residual become equal. The minimum-zone value never exceeds the peak-to-valley value from any other reference and is the smallest admissible form-error value.

Illustrative example

Schematic · not measured data
Schematic: containment bands of one profile under endpoint versus minimum-zone references, data illustrative.

Teaching data, not measured: five equally spaced heights [0, −0.5, +1.0, 0, +0.3] μm. The endpoint reference gives 1.5 μm; rotating the reference to a slope of 0.25 μm per station narrows the band to 1.25 μm with low-high-low alternating contacts. The example shows that the minimum-zone value is the smallest among references.

Limitations

  • The minimum-zone value holds only for the given point set; sparse sampling underestimates the error.
  • Without convergence and the alternating-contact criterion the band is not a true minimum zone.
  • Passing off a least-squares result scaled by an empirical factor as a minimum-zone value.

Reporting

Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.

Method reference ↗ Method reference ↗ 
STEP BY STEP

Understand the measurement step by step

Principle schematic · not to scale
1 / 6
  1. The 25 heights rise as a 5 × 5 field of bars, drawn one by one.
  2. A four-corner datum plane slides over the tops and the residual bars turn red and blue.
  3. The datum plane tilts slightly and the containment band closes like a clamp.
  4. The highest and lowest points touch the plane in turn and the contacts light up.
  5. The bandwidth read-out rolls from 5.76 μm down to 5.20 μm and locks.
  6. Contact lines and datum parameters are written into a report card.

Teaching animation: follow the stations and reference directions. Do not read dimensions or angles from the schematic.

WORKED EXAMPLE

Minimum-zone evaluation of a 25-point data set from a 400 mm × 400 mm plate

Teaching example with assumed data, not a customer project, instrument accuracy claim or acceptance result.

Example task

A diagonal-method run on a 400 mm × 400 mm plate has produced 25 relative heights on a 5 × 5 grid, referenced to the corner (0,0). The same data set is now evaluated twice, once against a four-corner datum and once by the minimum-zone criterion, to show how the choice of datum changes the reported flatness.

Setup and preparation

  • Arrange the 25 stations as a matrix with x=0/100/200/300/400 mm and y=0/100/200/300/400 mm, all heights in μm, the corner set to zero.

  • Check completeness: all 25 values valid, no gaps or duplicate numbers, maximum 3.82 μm and minimum −1.68 μm.

  • Fix the four-corner datum plane and compute the first containment band as a comparison baseline.

  • On a computer treat the two slopes (px, py) of the datum as variables and shrink the residual band with a golden-section search.

  • After convergence verify the contact pattern: maxima and minima must sit on opposite sides and alternate, and their counts and coordinates are recorded.

  • Record the optimal datum parameters, the minimum bandwidth and the contact coordinates, and report them together with the four-corner result.

Example data and records

  • Section y=0 mm (x=0/100/200/300/400): 0.00, -2.14, -2.68, -1.62, 1.04 μm

  • Section y=100 mm (x=0/100/200/300/400): 1.61, -0.46, -0.93, 0.20, 2.93 μm

  • Section y=200 mm (x=0/100/200/300/400): 2.22, 0.22, -0.18, 1.02, 3.82 μm

  • Section y=300 mm (x=0/100/200/300/400): 1.83, -0.10, -0.43, 0.84, 3.71 μm

  • Section y=400 mm (x=0/100/200/300/400): 0.44, -1.42, -1.68, -0.34, 2.60 μm

  • Instrument: autocollimator with 0.01″ resolution, reflector base span 100 mm, every step converted with Δh=L·tanθ

  • Environment: 20.1 °C, 0.2 °C drift over the run, windows closed and no foot traffic

  • Data are zeroed at the corner (0,0); negative values lie below that point

  • Three stations re-measured, difference from the first reading ≤0.04 μm

Calculation steps

Four-corner datum: the plane through (0,0), (400,0) and (0,400) has slopes px=0.00260 μm/mm and py=0.00110 μm/mm, equivalent to +0.23″ about X and +0.54″ about Y. The residuals are +0.00 → -2.40 → -3.20 → -2.40 → +0.00 → +1.50 → -0.83 → -1.56 → -0.69 → +1.78 → +2.00 → -0.26 → -0.92 → +0.02 → +2.56 → +1.50 → -0.69 → -1.28 → -0.27 → +2.34 → +0.00 → -2.12 → -2.64 → -1.56 → +1.12 μm, giving f₀=5.76 μm. Minimum-zone search: minimising max−min over (px, py) converges at px=0.00400 μm/mm and py=0.00250 μm/mm, a further (+0.29″, +0.29″) rotation relative to the four-corner datum. The upper extreme +1.72 μm touches at 2 points and the lower -3.48 μm at 2 points, on opposite sides of the datum, and the minimum-zone bandwidth is f_MZ=5.20 μm. The two datums differ by 0.56 μm, or 10.8% of the minimum-zone value, so on this mainly quadratic surface the datum choice matters.

Interpretation

Minimum-zone flatness 5.2 μm (four-corner datum gives 5.8 μm). Passes the 8 μm grade 1 tolerance; fails the 4 μm grade 0 tolerance by 1.2 μm.

Verification and pitfalls

  • The minimum-zone value must not exceed any other datum result: 5.20 μm ≤ 5.76 μm here, so the check passes.

  • Contacts must alternate: 2 points on the upper bound and 2 on the lower; a one-sided pattern means the search stopped in a local minimum.

  • Row and column indices of the matrix must match the measured coordinates, since a transposed matrix keeps the same bandwidth but moves the contacts.

  • Keep all heights in μm; feeding angles into the height matrix biases every result by three orders of magnitude.

  • Report the datum parameters and contact coordinates so a third party can reproduce the evaluation.