MICROCRE / OPTICAL METROLOGY400-812-7833  ·  SHANGHAI
研润 WHW
MICROCREPRECISION IN EVERY ANGLE
Home / Straightness of a 12 m gantry beam: rebuilding a profile from segment angles
CASE 01 · TEACHING DATA

Straightness of a 12 m gantry beam: rebuilding a profile from segment angles

Beam sag changes tool height along the travel. A long-range autocollimator reads the angle segment by segment at a 500 mm span, and accumulating those angles gives the height profile.

Object12 m gantry beam, one measuring line, 25 stations
InstrumentDual-axis autocollimator with 50 m range, reflector base with 500 mm span
AcceptanceFull-length straightness 0.030 mm (teaching assumption)
h_i = sum L tan(theta)

Where the difficulty lies

Across a 12 m path the spot grows and return light weakens, so air movement disturbs the reading. Every time the base is moved, span and orientation must stay identical.

Sag is a slowly varying shape. A single reading says nothing; only a continuous profile allows a decision about the middle of the beam.

Setup and acquisition

The autocollimator is fixed at one end of the beam facing the reflector. The reflector base advances 500 mm per step for 24 segments. Measure once in one direction, then re-check both ends in reverse to detect drift.

Record ambient temperature and the amplitude of reading wander. If wander exceeds roughly 0.3″, pause and repeat that segment once conditions settle.

Readings and calculation

The angle sequence rises monotonically from −2.2″ to +2.2″ and crosses zero in the middle, indicating a symmetric sag. Each increment follows Δh = 500 mm × tan θ, where 1″ corresponds to about 2.42 μm.

After accumulation the mid-span station sits about 33.4 μm below the end points. Referenced to the end-to-end line, the least-squares residual peak-to-valley is 33.4 μm.

Conclusion and action

The result of 0.0334 mm exceeds the 0.030 mm requirement by about 3.4 μm, so the beam fails. The symmetric shape points to stiffness or preload rather than local damage.

Adjust beam preload and gibs, then repeat on the same line with the same span. If the shape stays and the amplitude drops, the adjustment worked.

Case workflow, animated

ABCDΔ₄₁Closure c = ΣΔ should approach 0; c/4 is a conditional correctionSchematic · not to scale · not measured data
1 / 4
  1. Confirm the task and datum
  2. Lay out stations by span
  3. Acquire readings and convert
  4. Evaluate and verify

The animation follows the station layout and calculation order of this case. Values are teaching data.

Conclusion33.4 μm sag exceeds the 0.030 mm requirement; adjust preload and repeat under identical conditions.

The data are a teaching demonstration of the calculation. The conclusion applies to the static geometry of this line only, not to machining accuracy, and does not replace an acceptance report.

Recommended instruments