Principle
Mirror rotation changes the return-image position. Calibrated angular readings and the effective base span support height reconstruction.
Preparation
Define the line, travel, evaluation reference and tolerance. Clean the supports and verify reflector span, step spacing, calibration and thermal stability.
Workflow diagram
Procedure
- Mark equally spaced stations and record the actual span between the reflector base contacts.
- Align the autocollimator and verify the return beam over the full travel.
- Move the base in sequence and record stable angular readings without changing its orientation.
- Convert angular readings to height increments using the actual geometry; accumulate and evaluate using the agreed reference method.
- Repeat in reverse and report the profile, repeatability, environment and uncertainty information.
Calculation
θrad=θ″π/648000; Δh=L tan(θrad). For L in mm, multiply by 1000 for μm. Accumulate interval increments, then apply the specified evaluation reference.
Illustrative example
Illustration: L=100 mm and θ=1″ give about 0.4848 μm over one interval, not the straightness of the entire guideway.
Limitations
- An autocollimator measures reflector angle, not surface height directly.
- A mismatch between sampling steps and base span can invalidate reconstruction.
- One measured line does not establish the flatness of an entire surface.
Reporting
Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.