CASE 10 · TEACHING DATA

400 mm parting-line parallelism: judging margin behind 4.1″

Parting-line parallelism is usually quoted in millimetres, but what you measure is an angle. Converting 0.008 mm over 400 mm to 4.1″ shows how much margin is left.

Object400×300 mm injection mould, moving and fixed parting faces, five points
Instrument0.1″ autocollimator, 150 mm base, right-angle attachment
Conditions20±1 °C, mould rested 6 h, release agent removed
Criterionparallelism ≤0.010 mm over 400 mm, i.e. ≤5.2″
h_i = sum L tan(theta)

Get the unit conversion right

0.008 mm over 400 mm is 2×10⁻⁵ rad, i.e. 4.1″. The common field error is dividing an angle by 400 mm, mixing radians and degrees, which is off by a factor of 200.

Compare margins on the same reference length: if the drawing says 0.010 mm, do not compare against a value normalised to 1000 mm.

Can five points represent the face

Four corners plus centre catch overall tilt, but the real risk is local sink or residual plastic. When the result sits near the limit (4.1″ against 5.2″), densify to 3×3 or 4×4.

A sudden single-point deviation usually means residual plastic or a ding, not a machining error. Clean and re-measure before treating it as geometry.

Verification after closing

A good angle is not a good closure. After closing, check gap distribution with feeler gauges or pressure paper; a 0.02 mm hard spot causes flash.

If each half passes but the closed gap worsens, the guide pillars are not square to the parting face — go back to squareness, do not keep lapping the face.

Case workflow, animated

Image point (x, y)αX for pitch, Y for yawθx = x/2f, θy = y/2fSchematic · not to scale · not measured data
1 / 4
  1. Confirm the task and datum
  2. Lay out stations by span
  3. Acquire readings and convert
  4. Evaluate and verify

The animation follows the station layout and calculation order of this case. Values are teaching data.

Conclusion4.1″ (0.008 mm over 400 mm) is inside the 5.2″ limit with under 25% margin. Densify to 3×3 and confirm gap continuity with pressure paper after closing.

Teaching data. Real mould judgement also involves face width, vent placement and shrinkage. Parallelism passing does not mean the part passes.