Principle
The apex angle is measured by autocollimation: with the prism on a table, the autocollimator is autocollimated on each refracting face (return image on the reticle); the reading difference gives the angle φ between face normals and A=180°−φ. The deviation angle between incident and emergent rays is measured directly by rotating the autocollimator or the table. At minimum deviation the first-order sensitivity to incidence angle vanishes, making the measurement robust. With A and the minimum deviation δmin, the refractive index is n=sin[(A+δmin)/2]/sin(A/2).
Preparation
Clean the prism and inspect the transmitting faces for chips and scratches. Clamp it on the rotary table without stress. Match the autocollimator height to the table axis; the table resolution (e.g. 1″–10″ class) should pair with an autocollimator of 0.1″–0.5″ class. Record the ambient temperature.
Workflow diagram
Procedure
- Clamp the prism on the table so both refracting faces can be brought toward the autocollimator within rotation range.
- Rotate to autocollimate the first face (return image on the reticle) and record the table position φ1.
- Rotate to autocollimate the second face and record φ2; the apex angle A=180°−|φ2−φ1|.
- Re-measure both faces to confirm clamping stability and average the results.
- Observe the deviation in transmission: track the emergent beam versus table angle; the extremum is the minimum deviation δmin.
- Compute the refractive index if required and report A, δmin, n with conditions and uncertainty.
Calculation
Apex angle: A=180°−φ with φ=|φ2−φ1| the angle between face normals. Refractive index: n=sin[(A+δmin)/2]/sin(A/2). General deviation (teaching form): δ=i1+i2′−A with incidence i1 and emergence i2′; at minimum deviation dδ/di1=0. Conversion: 1″=π/648000 rad. The index uncertainty propagates from A and δmin via partial derivatives of the formula.
Illustrative example
Teaching data, not measured: autocollimation readings φ1=30°00′10″ and φ2=90°00′15″ give φ=60°00′05″ and A=59°59′55″. With δmin=40°00′20″, the demonstration index is about n≈sin(50°00′07.5″)/sin(29°59′57.5″)≈1.5322. No real prism is implied.
Limitations
- Clamping stress or temperature changes alter the apex angle; re-check between readings.
- Autocollimation targets the face normal, not the edge; writing φ instead of 180°−φ inverts the geometry.
- The deviation curve is flat near the minimum, so inadequate table resolution inflates the δmin uncertainty.
Reporting
Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.