Small Angles & Indexing Error
Compare small angular movements and precision indexing positions.
A polygon mirror with calibrated face angles plus an autocollimator gives full-circle indexing calibration.
Face normals have calibrated angular spacing. Aligning each face in turn compares the table reading with the nominal angle; the closure error checks consistency.
Standard numbers for checking method clauses. Full texts are not redistributed; listing does not certify an instrument.
Compare small angular movements and precision indexing positions.
Evaluate several rotary positions in both approach directions, separating position error, repeatability and reversal.
Compare angles between reflecting-face normals or deviations of polygon divisions.
Separate repeated-reading dispersion from changes caused by remounting or re-setup.
Verify index errors of rotary tables with a polygon and autocollimator.
Measure the apex angle by autocollimation, then derive or verify the deviation.
Convert the split half-field double image into the roof-angle error from 90°.
Verify the nominal 90° deviation of a pentaprism and its stability.
Derive the wedge from the angular separation of the two face reflections: α=Δθ/(2n).
A mirror on the spindle face tracks tilt over rotation for harmonic separation.
Quantify reading scatter from vibration via a time series on a static target.
Verify indication errors across the range with an angle generator or table and polygon.
Our catalogue focuses on angular metrology; this technique is listed for method comparison and system planning.