MICROCRE / OPTICAL METROLOGY400-812-7833  ·  SHANGHAI
研润 WHW
MICROCREPRECISION IN EVERY ANGLE
Home / Repeatability & Re-setup Verification
MEASUREMENT METHOD

Repeatability & Re-setup Verification

Separate repeated-reading dispersion from changes caused by remounting or re-setup.

Principle

Repeatability describes dispersion under stated conditions; re-setup adds mounting and datum-realization effects.

Preparation

Fix the station, sampling, settling criterion and repetition count; separate unchanged-setup and remounting trials.

Workflow diagram

01Collect repeated readings without changing the setup02Record times03Remount or reposition according to a separate trial pl04Evaluate within-group dispersion and between-group mea05Report sample size

Procedure

  1. Collect repeated readings without changing the setup.
  2. Record times, environment and abnormal events.
  3. Remount or reposition according to a separate trial plan.
  4. Evaluate within-group dispersion and between-group mean differences separately.
  5. Report sample size, conditions and statistics explicitly.

Calculation

Sample standard deviation s=sqrt[Σ(xᵢ−x̄)²/(n−1)]. For independent identically distributed readings, the Type A standard uncertainty of the mean is uA=s/√n.

Illustrative example

Schematic · not measured data
Principle and workflow schematic. Numbers are worked examples, not customer or instrument test results.

Illustration: for ten independent readings with s=0.20″, uA of the mean is about 0.063″, excluding calibration and other sources.

Limitations

  • Continuous samples may be autocorrelated.
  • Good repeatability does not prove absence of systematic error.

Reporting

Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.

Method reference ↗ Method reference ↗ Method reference ↗ 
STEP BY STEP

Understand the measurement step by step

x̄12345Repeated readings scatter about the mean; not a drift curveReading numberPrinciple schematic · not to scale
1 / 4
  1. Repeat fixed setup · Fix the target and setup, define a settling time and plan five readings without remounting.
  2. Compute within-group statistics · Record the first illustrative group and compute its mean and sample standard deviation; close the group before any mounting change.
  3. Form remounting groups · Start a separate remounting record with a mean for each new placement; keep group identifiers rather than appending means to raw observations.
  4. Interpret separately · Show within-group scatter and shifts between group means; a proper variance-component analysis requires balanced replicated groups and a suitable model.

Teaching animation: follow the stations and reference directions. Do not read dimensions or angles from the schematic.

WORKED EXAMPLE

Separating fixed-setup repeats from remounting groups

Teaching example with assumed data, not a customer project, instrument accuracy claim or acceptance result.

Example task

Teaching demonstration, not a real customer project: five fixed-setup observations illustrate short-term repeatability, while three remounting-group means are analysed separately. Different variation sources are not combined into one number.

Setup and preparation

  • Fix the target and setup, define a settling time and plan five readings without remounting.

  • Record the first illustrative group and compute its mean and sample standard deviation; close the group before any mounting change.

  • Start a separate remounting record with a mean for each new placement; keep group identifiers rather than appending means to raw observations.

  • Show within-group scatter and shifts between group means; a proper variance-component analysis requires balanced replicated groups and a suitable model.

Example data and records

  • Fixed-setup observations: [1.0,1.1,0.9,1.0,1.0] arcsec, n=5.

  • Group mean 1.0 arcsec; sample standard deviation 0.070711 arcsec.

  • Illustrative remounting-group means: [1.00,1.25,0.85] arcsec; their sample standard deviation is 0.202073 arcsec.

Calculation steps

Within-group s=√[0.02/(5−1)]=0.070711 arcsec; with independent identically distributed readings, uA(mean)=s/√5=0.031623 arcsec. The 0.202073 arcsec standard deviation of group means is a different statistic, not an isolated mounting variance component.

Interpretation

Short-term observations and remounting means show different scatter. This is grouped arithmetic, not a total-accuracy claim or formal repeatability acceptance test.

Verification and pitfalls

  • Drift or autocorrelation can invalidate the independence assumption behind s/√n.

  • Record group size, operator, mounting and environmental changes.

  • Use a sufficiently replicated hierarchical design to separate variance components.