Small Angles & Indexing Error
Compare small angular movements and precision indexing positions.
Uses a higher-grade encoder or polygon as reference to evaluate full-circle error of tables and indexing mechanisms.
The tested axis and reference encoder turn together; synchronous angle readings give a difference series whose harmonics separate eccentricity, graduation and random components.
Standard numbers for checking method clauses. Full texts are not redistributed; listing does not certify an instrument.
Compare small angular movements and precision indexing positions.
Evaluate several rotary positions in both approach directions, separating position error, repeatability and reversal.
Separate repeated-reading dispersion from changes caused by remounting or re-setup.
Separate offset from tilt using angular readings in two sections.
Repeatable stations turn single measurements into a trend you can act on.
Verify index errors of rotary tables with a polygon and autocollimator.
Use angle gauges as material standards and read the difference with an autocollimator.
A mirror on the spindle face tracks tilt over rotation for harmonic separation.
Quantify reading scatter from vibration via a time series on a static target.
Our catalogue focuses on angular metrology; this technique is listed for method comparison and system planning.