Principle
The reference supplies the large angle; the instrument reads a small residual within range.
Preparation
Select a calibrated polygon or angular generator; define positions, cycles, settling and direction.
Workflow diagram
Procedure
- Verify stable mounting of the rotary stage and reference.
- Approach each position forward and record residual small angles.
- Repeat in reverse, preserving the approach-direction record.
- Apply reference corrections and obtain position deviations.
- Evaluate directional means, dispersion and reversal at each position.
Calculation
At each position, Bᵢ=mean(eᵢ,forward)−mean(eᵢ,reverse). Retain raw values and dispersion separately; use the specified standard for formal metrics.
Illustrative example
Illustration: mean deviations +0.6″ forward and +0.2″ reverse give +0.4″ directional difference.
Limitations
- One cycle or one direction cannot separate all repeatability and reversal effects.
- An uncalibrated polygon is not an error-free reference.
Reporting
Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.