Principle
An angular reference generates the nominal rotation; the autocollimator measures the remaining small difference.
Preparation
Define the angular range, calibration corrections, measurement axes, target positions and approach directions.
Workflow diagram
Procedure
- Fix the test item and angular reference and check the beam return and mounting stability.
- Set the reference reading and confirm the instrument units and angular convention.
- Move through the specified angular positions and record the direction of approach.
- Compare readings with reference angles, applying calibration corrections and repeating the sequence.
- Report indexing error, repeatability and reversal effects separately.
Calculation
Error is actual angle minus nominal angle. Apply reference-certificate corrections with signs defined by the measurement model; process both approach directions separately.
Illustrative example
Illustration: a corrected position of 10°+0.4″ has a +0.4″ error relative to nominal 10°.
Limitations
- Resolution is not measurement accuracy.
- Displays calibrated in mirror angle must not be divided by two again because the beam is reflected.
- Range, cross-axis sensitivity and reference-surface quality affect results.
Reporting
Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.