MEASUREMENT METHOD

Calibrating Autocollimator Indication Errors

Determine the indication error of an autocollimator over its full range, including the working interval, yielding point corrections and calibration uncertainty. It serves acceptance, periodic calibration and post-repair verification. Indication error propagates directly into every downstream arcsecond result.

Principle

The calibration feeds the autocollimator known angles and compares indications with the standards. Two standard families are common: small-angle generators (flexure-based sine mechanisms or interferometric angle rigs) producing precisely known angles near zero, or a polygon on a high-accuracy rotary table producing standard rotations around the circle. A mirror on the moving part of the standard carries the angles while the autocollimator reads them; the differences are the indication errors. The range is sampled at equal steps (e.g. every 100″ or 1′ depending on range and class) in both directions to assess hysteresis and drift.

Preparation

Confirm the standard's traceability and uncertainty, which should be better than the unit under test, typically by a factor of three or more (rule of thumb). Mount the mirror on the moving part of the standard, normal to the autocollimator. Warm up the instrument for the specified time and record temperature. Plan the calibration points to cover the range and the working interval.

Workflow diagram

01Choose the standard02Mount mirror and warm up03Calibrate forward04Calibrate reverse05Errors and corrections06Issue calibration report

Procedure

  1. Set up the standard with confirmed traceability; mount the mirror on the moving part and align it to the autocollimator.
  2. Warm up the instrument, record the zero reading and begin.
  3. Present the standard angles point by point as planned, recording the autocollimator indication.
  4. Repeat the points in reverse after completing the forward run to check hysteresis and zero drift.
  5. Compute indication errors e=indication−standard angle and, if needed, fit an error curve for interpolated corrections.
  6. Report per-point errors or corrections, forward-reverse agreement, zero drift and calibration uncertainty.

Calculation

Per-point error e_i=R_i−S_i, with R_i the indication and S_i the standard angle. The forward-reverse difference h_i=|e_i(fwd)−e_i(rev)| indicates hysteresis. The correction is c_i=−e_i, added to indications in use. Zero drift d=final−initial zero reading; data from out-of-tolerance intervals are rejected or re-measured. A combined uncertainty is u(e)≈√(u_S²+u_R²), u_S from the standard and u_R the repeatability (simplified example).

Illustrative example

Schematic · not measured data
Schematic: point-by-point comparison of standard angles and indications; values illustrative.

Teaching data, not measured: at the +500″ point an autocollimator shows errors of +0.4″ forward and +0.2″ reverse (demonstration), with a standard uncertainty of 0.1″ (k=1, example). The correction at that point is −0.3″ using the mean of both runs; the report also includes the 0.2″ hysteresis and zero drift. No real instrument is implied.

Limitations

  • Calibrating only near zero leaves the working interval's errors uncovered.
  • A standard with uncertainty comparable to the unit under test voids the calibration.
  • Unrecorded hysteresis and zero drift make corrections ill-suited for reverse travel.

Reporting

Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.

Method reference ↗ Method reference ↗ 
STEP BY STEP

Understand the measurement step by step

Principle schematic · not to scale
1 / 6
  1. The generator faces the autocollimator with the reflector bonded to its moving member.
  2. The generator steps in 60″ increments while the standard and indication are shown side by side and the error column fills.
  3. After +300″ the travel reverses and a small hysteresis loop appears in the error curve.
  4. Three repeats at +180″ show 0.01″ on the repeatability bar chart.
  5. The error curve is drawn against the ±0.1″ limit band with the 0.07″ point marked in red.
  6. A certificate card is produced with error, hysteresis, drift and the ppm correction.

Teaching animation: follow the stations and reference directions. Do not read dimensions or angles from the schematic.

WORKED EXAMPLE

Calibration of a 0.01″ autocollimator over ±300″

Teaching example with assumed data, not a customer project, instrument accuracy claim or acceptance result.

Example task

A periodic calibration is performed on an autocollimator with 0.01″ resolution over a ±300″ range. The reference is a sine-bar small-angle generator (0.005″ resolution, 0.02″ certificate uncertainty, k=2) with the reflector fixed to its moving member. Points are taken at ±60″, ±120″, ±180″, ±240″ and ±300″ in both directions to evaluate indication error, hysteresis and zero drift.

Setup and preparation

  • Place the generator on an isolation table, fix the reflector to its moving member and aim it at the autocollimator within 5′ of normal incidence.

  • Warm the instrument up for 60 min, wait for a stable reading and record the zero indication, +0.00″.

  • Step the reference in 60″ increments, wait 10 s per point and record the indication while travelling forward to +300″.

  • Return point by point and record the indications from −60″ to −300″ for the hysteresis evaluation.

  • Repeat three readings at +180″ for repeatability, then read the zero again at the end of the run.

  • 20.0 °C with drift under 0.3 °C; the reference certificate is within its validity period.

Example data and records

  • Reference: sine-bar small-angle generator, resolution 0.005″, certificate uncertainty 0.02″ (k=2)

  • Zero: +0.00″ after 60 min warm-up, +0.02″ at the end of the run

  • +60″: indication +60.02″, error +0.02″

  • +120″: indication +120.05″, error +0.05″

  • +180″: indication +179.96″, error -0.04″

  • +240″: indication +240.03″, error +0.03″

  • +300″: indication +300.07″, error +0.07″

  • −60″: indication -59.97″, error +0.03″

  • −120″: indication -119.96″, error +0.04″

  • −180″: indication -180.05″, error -0.05″

  • −240″: indication -239.99″, error +0.01″

  • −300″: indication -299.95″, error +0.05″

  • Three repeats at one standard angle: errors 0.02, 0.03, 0.02″, repeatability 0.01″

Calculation steps

The indication error is e=indication−standard. Forward: +0.02 → +0.05 → -0.04 → +0.03 → +0.07 ″; reverse: +0.03 → +0.04 → -0.05 → +0.01 → +0.05 ″. The largest error over the range is |e|max=0.07″ at +300″. Hysteresis, the difference between forward and reverse errors at the same point, is +0.01 → +0.01 → +0.01 → +0.02 → +0.02 ″ with a maximum of 0.02″. The sensitivity deviation from the two range ends is (e(+300)+e(−300))/2 ÷ 300 = (0.07+(0.05))/2/300 = 2.00e-04, i.e. 200 ppm, which can be applied as a proportional correction in software. Zero drift is 0.02″ and repeatability is 0.01″. The combined uncertainty u=√(0.02²+0.01²+0.01²)≈0.024″ is far below the largest error 0.07″, so the error verdict stands.

Interpretation

Largest indication error 0.07″, largest hysteresis 0.02″ and zero drift 0.02″ are all inside the ±0.1″ MPE, so the instrument passes for continued use over ±300″; a proportional correction of 200 ppm is recommended in software.

Verification and pitfalls

  • Calibration points must cover the full range in both signs: forward-only data would miss the 0.02″ hysteresis and the zero asymmetry.

  • The reference certificate must be valid and its uncertainty below one third of the instrument MPE (0.02″ against 0.1″ here).

  • Allow at least 10 s settling per point; the generator springs back slightly after indexing and an immediate reading is 0.03″ off.

  • Read the zero before and after the run; a drift above 0.05″ means the whole data set must be repeated.

  • The bond between reflector and moving member must be stiff; a soft adhesive adds more than 0.05″ of random error.