Home / 1500×1000 granite plate: a 9×7 grid and two lapping passes
CASE 06 · TEACHING DATA

1500×1000 granite plate: a 9×7 grid and two lapping passes

Out-of-flatness on a large plate usually concentrates at a few high points. A 9×7 grid locates them, a planned lapping path removes them, and a re-measure confirms convergence.

Object1500×1000 mm granite plate, 9×7 grid, 187.5 / 166.7 mm pitch
Instrument0.1″ autocollimator, 100 mm base, 90° folding mirror
Conditions20±0.5 °C, 24 h soak, three-point support
CriterionGB/T 20428, grade 000 allows 3.5 μm
h_i = sum L tan(theta)

Laying out the grid

Sixty-three points sounds many, but on a large plate a single diagonal can miss a local bowl or saddle. A 9×7 grid resolves features around 190 mm, which matches the lap diameter.

The grid must be parallel to the edges with a half-pitch margin so chamfers do not affect readings. Number the points and re-measure the same numbers after lapping.

From height map to lapping path

Integrating slopes into heights and fitting a plane leaves +1.3 μm to −1.6 μm, peak-to-valley 2.9 μm, RMS 0.9 μm. High points cluster at upper left and lower right, diagonally.

Do not aim for one pass: each round targets points more than 0.5 μm high with short, cross-direction strokes, removing 1–2 μm per pass. Rushing leaves a new central crown.

Judging convergence

Pass one took 5.6 μm to 3.9 μm, pass two to 2.9 μm, and pass three improved only 0.2 μm — stop and measure in the metrology room rather than chase further.

Re-soak after every pass. Lapping heat expands the plate locally by about 1 μm, and hot readings mislead the next path.

Case workflow, animated

Minimum zone: two parallel planesLeast squares: minimal residual sumDifferent results — state the method in the reportSchematic · not to scale · not measured data
1 / 4
  1. Confirm the task and datum
  2. Lay out stations by span
  3. Acquire readings and convert
  4. Evaluate and verify

The animation follows the station layout and calculation order of this case. Values are teaching data.

ConclusionTwo passes brought peak-to-valley from 5.6 μm to 2.9 μm, inside the 3.5 μm grade 000 limit. A third pass improved only 0.2 μm, so stop and move to final measurement.

Teaching data. Real plate repair also depends on support scheme, vibration and lap wear. Peak-to-valley is not minimum-zone flatness; formal judgement needs the standard algorithm.