Home / Optics & Optomechanics
APPLICATION GUIDE

Optics & Optomechanics

Mirror assemblies, optical tables, rotary stages and imaging-system mounts.

Methods

Procedure

  1. Define optical datums
  2. Align the path
  3. Measure angles
  4. Adjust mounting
  5. Recheck stability

Angular alignment alone does not establish wavefront quality or imaging performance.

MEASUREMENT SCENARIOS

Typical measurement scenarios in this industry

Each scenario gives the object, tolerance range and method, ready to take to the selector.

Open in the atlas ↗
AssemblyOptical Alignment

Mirror assembly optical axis alignment

After the mirror sits in its cell, watch the returned image and adjust pitch and yaw screws into 5–30″; lock and re-check drift.

Wide-field electronic autocollimators↗
InspectionOptical Alignment

Lens decentre and group coaxially

Rotate the group and watch the image wobble; the radius of the traced circle gives decentre. Judge element by element so error does not accumulate in the assembly.

Wide-field electronic autocollimators↗
AssemblyStage Pitch & Yaw

Optical table levelling and post-isolation stability

Five points (four corners plus centre), level to within 10″, then log for 30 min with air isolation on to confirm no slow tilt.

Metrology autocollimator series↗
InspectionPrism & Polygon Angle Comparison

Prism and wedge angle verification

90° prisms and small wedge angles. Use a polygon or angle gauge, measure face by face, and correct for incidence angle and refractive index.

Goniometers↗
InspectionSmall Angles & Indexing Error

Substrate figure change after coating

Coating stress bows the substrate. Compare the reflected angle at the same point before and after coating to decide whether tooling compensation is needed.

Zhongji automatic measurement series↗
AssemblyOptical Alignment

Interferometer path alignment and reference flat attitude

Reference flat attitude relative to the stage affects fringe reading; monitor alignment with an autocollimator down to arc-seconds.

General-purpose electronic autocollimators↗
AssemblyOptical Alignment

Lens centring and axis consistency

Groups 5–200 mm, centring 5–20 μm; rotation gives image jump.

General-purpose electronic autocollimators↗
InspectionPrism & Polygon Angle Comparison

Prism and wedge angle

10°–90°, typically 2–10″; autocollimator with a rotary table and closure check.

Goniometers↗

Recommended instruments

STEP BY STEP

Understand the measurement step by step

  1. 01

    Define optical datums

  2. 02

    Align the path

  3. 03

    Measure angles

  4. 04

    Adjust mounting

  5. 05

    Recheck stability

d₁d₂θx_midLSection 1Section 2Two sections separate position offset from angular tiltPrinciple schematic · not to scale
1 / 4
  1. Set two planes · Define one mechanical reference axis and install near and far reference planes 600 mm apart, checking all distance units.
  2. Locate beam centres · Without changing the beam state, record its centre at both planes using consistent coordinate signs and the same centre-detection rule.
  3. Separate tilt and offset · Calculate the directional slope from the position difference and use arctangent for tilt. Record near-plane decentring separately.
  4. Verify after locking · Use the available assembly adjustments to address translation and tilt separately. After locking, remeasure both planes and check centre-detection repeatability.

Teaching animation: follow the stations and reference directions. Do not read dimensions or angles from the schematic.

WORKED EXAMPLE

Optomechanical assembly: separate decentring from tilt using two planes

Teaching example with assumed data, not a customer project, instrument accuracy claim or acceptance result.

Example task

Teaching demonstration, not a real customer project. Beam-centre positions along one transverse coordinate are read at two reference planes separated by a known distance. A single-plane position cannot independently determine axis tilt. Position readings require a suitable target or position-measurement system.

Setup and preparation

  • Define one mechanical reference axis and install near and far reference planes 600 mm apart, checking all distance units.

  • Without changing the beam state, record its centre at both planes using consistent coordinate signs and the same centre-detection rule.

  • Calculate the directional slope from the position difference and use arctangent for tilt. Record near-plane decentring separately.

  • Use the available assembly adjustments to address translation and tilt separately. After locking, remeasure both planes and check centre-detection repeatability.

Example data and records

  • Near-plane transverse offset d1=+20 μm.

  • Far-plane offset d2=+50 μm; plane separation L=600 mm.

  • Only one transverse coordinate is illustrated; repeat independently for the other.

Calculation steps

Slope=(50−20)/(600×1000)=0.00005. θ=atan(0.00005)≈50 μrad=10.313240″. Near-plane decentring remains +20 μm.

Interpretation

The illustrated axis tilt is approximately 10.313240″ in this direction, with +20 μm near-plane decentring. These are distinct quantities; one angular adjustment does not guarantee that both vanish.

Verification and pitfalls

  • Use a common mechanical datum and consistent coordinate direction at both planes.

  • Spot positions are not autocollimator angle readings; this example makes no product position-measurement claim.

  • The example does not evaluate wavefront, imaging quality, focal length or overall optical performance.