Home / Endpoint-Reference Evaluation
MEASUREMENT METHOD

Endpoint-Reference Evaluation

Provide a simple, manually checkable reference for reconstructed straightness profiles, common in in-process checks and quick screening. For approximately convex or concave profiles the endpoint result approaches the minimum-zone value. It also supports comparisons under a fixed-reference convention.

Principle

A straight line joins the first and last measured points and serves as the evaluation reference. Every station deviation is computed against this line, the two endpoints having zero deviation. The error is the sum of the largest positive and the magnitude of the largest negative deviation; if all deviations share one sign, the largest absolute deviation is taken. The reference is not rotated, so the result generally equals or exceeds the minimum-zone value, but it is transparent and easy to verify.

Preparation

Confirm station spacing and count, and check the quality of the first and last readings, since these define the reference and faults there propagate globally. Keep the raw integrated profile without smoothing. Label the reference method in the report to avoid confusion with other evaluations.

Workflow diagram

01Integrate profile02Join endpoints03Compute deviations04Take extremes05Sum and report06Report and method label

Procedure

  1. Integrate the raw angular record into station heights, setting the first point to zero.
  2. Define the reference line from the endpoints: zref,i=(zi−z1)·(xi−x1)/(xn−x1)+z1.
  3. Compute deviations di=zi−zref,i; the endpoints must give zero, serving as a check.
  4. Identify the largest positive deviation dmax+ and the largest negative deviation dmax−.
  5. The profile error is f=|dmax+|+|dmax−| (the largest absolute value if all share one sign).
  6. Report the value, reference method and raw data, noting that it is not a minimum-zone result.

Calculation

The two-point reference is zref(x)=z1+(zn−z1)·(x−x1)/(xn−x1). Deviations are di=zi−zref(xi). The error f=max(di)+|min(di)|, or f=max|di| when all deviations share one sign. It always equals or exceeds the minimum-zone value, an upper bound of the form error.

Illustrative example

Schematic · not measured data
Schematic: deviations referenced to the endpoint line; values are illustrative.

Teaching data, not measured: a 400 mm line, five equally spaced points, heights [0, +0.8, +1.6, +0.8, 0] μm. The reference is the x-axis, deviations share one sign and f=1.6 μm. With heights [0, −0.5, +1.0, −0.5, 0] μm, f=|+1.0|+|−0.5|=1.5 μm. The numbers only demonstrate the arithmetic.

Limitations

  • Faulty first or last readings tilt the whole reference and distort the error.
  • For S-shaped profiles the endpoint result markedly exceeds the minimum zone and may falsely fail a part.
  • Values from different references are not comparable; the report must state the method.

Reporting

Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.

Method reference ↗ Method reference ↗ 
STEP BY STEP

Understand the measurement step by step

Principle schematic · not to scale
1 / 5
  1. Side view of the bed with 11 stations marked as the base advances from tailstock to headstock.
  2. Each reading accumulates into a polyline that rises and then falls back.
  3. A dashed datum line is drawn between the two end points and the polyline is straightened onto it.
  4. The residual curve appears with the highest and lowest stations labelled.
  5. Two parallel lines clamp the residual curve and show 6.0 μm.

Teaching animation: follow the stations and reference directions. Do not read dimensions or angles from the schematic.

WORKED EXAMPLE

Straightness of a 2000 mm lathe bed in the vertical plane (end-point line)

Teaching example with assumed data, not a customer project, instrument accuracy claim or acceptance result.

Example task

A C6140 lathe bed is re-checked after hand scraping. The guideway is 2000 mm long and its straightness is evaluated in the vertical plane. The autocollimator stands on a rigid bracket at the tail end, the reflector base span is 200 mm, and the run covers 10 segments and 11 stations. The shop is at 22.6 °C and the bed has been levelled and left to settle for 8 h.

Setup and preparation

  • Clean the guideway, stone off scraping burrs and mark 11 stations at a 200 mm pitch.

  • Bolt the autocollimator to the tail bracket so the optical axis is nearly parallel to the guideway in the vertical plane, with residual pitch below 20′.

  • Seat both feet of the reflector base on the same generatrix and step it heel-to-toe; never let the base straddle an oil hole or relief groove.

  • Warm up for 20 min, zero the instrument and make a dry run to confirm that no reading exceeds the ±300″ range.

  • Record 22.6 °C and keep the shop door closed to avoid draughts and sudden temperature changes.

Example data and records

  • Segment 1 (0→200 mm): +1.20″

  • Segment 2 (200→400 mm): +2.00″

  • Segment 3 (400→600 mm): +1.40″

  • Segment 4 (600→800 mm): +0.30″

  • Segment 5 (800→1000 mm): -0.80″

  • Segment 6 (1000→1200 mm): -1.60″

  • Segment 7 (1200→1400 mm): -2.20″

  • Segment 8 (1400→1600 mm): -1.40″

  • Segment 9 (1600→1800 mm): -0.50″

  • Segment 10 (1800→2000 mm): +0.90″

  • Station 1 (0 mm) zeroed; station 11 (2000 mm) closure -0.68 μm

Calculation steps

Scale factor k=200×π/648000×1000=0.970 μm per arcsec, giving height steps Δh=+1.16 → +1.94 → +1.36 → +0.29 → -0.78 → -1.55 → -2.13 → -1.36 → -0.48 → +0.87 μm. Cumulating from the first station: +0.00 → +1.16 → +3.10 → +4.46 → +4.75 → +3.98 → +2.42 → +0.29 → -1.07 → -1.55 → -0.68 μm. End-point datum: the end-to-end rise is -0.68 μm, shared over 10 segments, so each segment is corrected by -0.068 μm. Removing the linear term leaves the residuals dᵢ=+0.00 → +1.23 → +3.24 → +4.66 → +5.02 → +4.31 → +2.83 → +0.77 → -0.52 → -0.94 → +0.00 μm, with exactly zero at both ends as the definition requires. The largest positive deviation +5.02 μm is at station 5 (800 mm) and the largest negative -0.94 μm at station 10 (1800 mm). The end-point straightness is f=5.02−(-0.94)=5.96 μm, equal to 3.0 μm/m or an equivalent tilt of 0.6″.

Interpretation

Straightness 6.0 μm against the end-point line over the full 2000 mm. This passes the 0.015 mm full-length tolerance for a lathe bed with 9.0 μm margin.

Verification and pitfalls

  • The datum is the straight line between the first and last station: both residuals read 0.00 μm and the -0.068 μm per-segment correction can be recomputed by hand.

  • This method never rotates the datum, so the result cannot be smaller than the minimum-zone value; a smaller number means a sign or correction error.

  • Repeat readings at one station must agree within 0.1″; otherwise check whether a foot sits on an oil hole or a burr.

  • Re-measure the first and last stations; if the cumulative closure exceeds 0.3 μm, repeat the whole profile.

  • Never adjust the bed shims mid-run: any adjustment invalidates every station already measured.