Principle
A straight line joins the first and last measured points and serves as the evaluation reference. Every station deviation is computed against this line, the two endpoints having zero deviation. The error is the sum of the largest positive and the magnitude of the largest negative deviation; if all deviations share one sign, the largest absolute deviation is taken. The reference is not rotated, so the result generally equals or exceeds the minimum-zone value, but it is transparent and easy to verify.
Preparation
Confirm station spacing and count, and check the quality of the first and last readings, since these define the reference and faults there propagate globally. Keep the raw integrated profile without smoothing. Label the reference method in the report to avoid confusion with other evaluations.
Workflow diagram
Procedure
- Integrate the raw angular record into station heights, setting the first point to zero.
- Define the reference line from the endpoints: zref,i=(zi−z1)·(xi−x1)/(xn−x1)+z1.
- Compute deviations di=zi−zref,i; the endpoints must give zero, serving as a check.
- Identify the largest positive deviation dmax+ and the largest negative deviation dmax−.
- The profile error is f=|dmax+|+|dmax−| (the largest absolute value if all share one sign).
- Report the value, reference method and raw data, noting that it is not a minimum-zone result.
Calculation
The two-point reference is zref(x)=z1+(zn−z1)·(x−x1)/(xn−x1). Deviations are di=zi−zref(xi). The error f=max(di)+|min(di)|, or f=max|di| when all deviations share one sign. It always equals or exceeds the minimum-zone value, an upper bound of the form error.
Illustrative example
Teaching data, not measured: a 400 mm line, five equally spaced points, heights [0, +0.8, +1.6, +0.8, 0] μm. The reference is the x-axis, deviations share one sign and f=1.6 μm. With heights [0, −0.5, +1.0, −0.5, 0] μm, f=|+1.0|+|−0.5|=1.5 μm. The numbers only demonstrate the arithmetic.
Limitations
- Faulty first or last readings tilt the whole reference and distort the error.
- For S-shaped profiles the endpoint result markedly exceeds the minimum zone and may falsely fail a part.
- Values from different references are not comparable; the report must state the method.
Reporting
Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.