Verification plan for a CMM granite table
For periodic verification and post-repair re-checks of a CMM granite table (1000×800 mm, example/teaching data). Autocollimator diagonal and grid lines are primary; an electronic level cross-checks shared lines. Confirm the support state and evaluation datum first, then measure, unify the datum and evaluate, run a reverse check and deliver verification records.
Instruments
- Photoelectric autocollimator (0.1″ class) with 150 mm reflector base (example)
- Calibrated electronic level
- Means to confirm the three-point support per drawing
- Temperature and station marking tools
Procedure
- Confirm the verification area 1000×800 mm (example), support positions per drawing, line layout and evaluation datum. Without an agreed tolerance, report data only.
- Clean the table and lay out two diagonals plus a longitudinal/transverse grid with a 150 mm span; mark stations and log temperature.
- Read each line with the autocollimator to 0.1″; the electronic level reads two shared lines simultaneously for cross-checking.
- Integrate angles into heights via Δh=L·tanθ and tie all lines to a common datum through shared nodes, forming the height matrix.
- Compare the two instruments on shared lines; if the rms height difference exceeds the agreed limit (2 μm, example), check supports, temperature and instrument state.
- Evaluate flatness against the agreed datum and reverse-repeat one diagonal.
- Deliver verification records: layout, raw data, height matrix, comparison log, flatness conclusion and traceability information.
Calculation
Δh=L·tanθ; with L=150 mm, 1″≈0.727 μm (example/teaching data). Instrument agreement is expressed as the rms height difference on shared lines. Flatness is computed on the agreed datum (minimum zone or least squares) and explicitly named.
Reporting
- Support state matches the drawing and is recorded
- Sign conventions and spans of both instruments unified
- Closure at shared nodes and inter-instrument differences checked
Limitations
Does not cover long-term granite ageing stability or the CMM probing error. The verification interval, tolerance and verdict follow the user's metrology specification or contract; this plan provides geometric data and comparison records only.
Understand the measurement step by step
- 01
Confirm the verification area 1000×800 mm (example), support positions per drawing, line layout and evaluation datum. Without an agreed tolerance, report data only.
- 02
Clean the table and lay out two diagonals plus a longitudinal/transverse grid with a 150 mm span; mark stations and log temperature.
- 03
Read each line with the autocollimator to 0.1″; the electronic level reads two shared lines simultaneously for cross-checking.
- 04
Integrate angles into heights via Δh=L·tanθ and tie all lines to a common datum through shared nodes, forming the height matrix.
- 05
Compare the two instruments on shared lines; if the rms height difference exceeds the agreed limit (2 μm, example), check supports, temperature and instrument state.
- 06
Evaluate flatness against the agreed datum and reverse-repeat one diagonal.
- 07
Deliver verification records: layout, raw data, height matrix, comparison log, flatness conclusion and traceability information.
- Thermal soak · Table cleaned and soaked at 20 °C, temperature logged.
- Segment setup · The 200 mm bridge runs end-to-end along the diagonal.
- Acquire readings · Read the angle per segment, averaging two observations.
- Convert heights · Convert arc-second readings into micrometre heights.
- Cumulate & evaluate · Cumulate heights and derive peak-to-valley straightness.
Teaching animation: follow the stations and reference directions. Do not read dimensions or angles from the schematic.
CMM Granite Table Verification · Tutorial Case
Teaching example with assumed data, not a customer project, instrument accuracy claim or acceptance result.
Example task
Teaching demonstration, not a real customer project: segmented angular measurement along a diagonal of a CMM granite table in a metrology room; data are instructional settings.
Setup and preparation
Stabilise the table at 20 °C±0.5 and clean the surface
Set up the autocollimator and a 200 mm bridge along the diagonal
Measure segment by segment from one corner along the diagonal
Take two readings per segment and average them
After three segments, evaluate straightness by the cumulative method
Example data and records
Bridge span L=200 mm; segment 1 autocollimator reading θ₁=+0.5″
Segment 2 autocollimator reading θ₂=-0.3″
Segment 3 autocollimator reading θ₃=+1″
Calculation steps
Height conversion: Δh = L·tan(θ·π/648000)×1000 (μm). Segment 1: 200×tan(+0.5″)=+0.48 μm; Segment 2: 200×tan(-0.3″)=-0.29 μm; Segment 3: 200×tan(+1″)=+0.97 μm. Cumulative heights: 0→+0.48→+0.19→+1.16 μm; peak-to-valley PV=+1.16−(+0.00)=+1.16 μm.
Interpretation
Tutorial result PV≈1.16 μm, demonstrating micrometre-level verification of a granite table straightness against teaching criteria.
Verification and pitfalls
Bridge support points rest on sound granite, clear of pores
Difference between repeat readings must not exceed 0.2″
Check instrument zero drift; return-to-zero error within 0.1″