In-house angular calibration capability
Build internal calibration for autocollimators, electronic levels and angle standards, including equipment, environment and uncertainty budget.
Instruments
- Reference autocollimator and small-angle tester
- Polygon or angle blocks
- Temperature-controlled room and isolation
Procedure
- Define items, range and target uncertainty.
- Select standards and support equipment along the traceability chain.
- Set up environmental monitoring, equipment files and interim checks.
- Write the procedure, record template and uncertainty budget.
- Run repeatability and comparison tests to confirm capability.
Calculation
The budget covers standard, repeatability, eccentricity, environment and resolution; combine in quadrature and apply the coverage factor.
Reporting
- Traceability complete
- Continuous environment records
- Budget reproducible
- Interim check criteria
Limitations
Formal accreditation follows local metrology authority requirements; site content does not replace assessment.
Understand the measurement step by step
- 01
Define items, range and target uncertainty.
- 02
Select standards and support equipment along the traceability chain.
- 03
Set up environmental monitoring, equipment files and interim checks.
- 04
Write the procedure, record template and uncertainty budget.
- 05
Run repeatability and comparison tests to confirm capability.
- Step 1 · Define items, range and target uncertainty.
- Step 2 · Select standards and support equipment along the traceability chain.
- Step 3 · Set up environmental monitoring, equipment files and interim checks.
- Step 4 · Write the procedure, record template and uncertainty budget.
- Step 5 · Run repeatability and comparison tests to confirm capability.
Teaching animation: follow the stations and reference directions. Do not read dimensions or angles from the schematic.
Worked uncertainty budget for an angle standard
Teaching example with assumed data, not a customer project, instrument accuracy claim or acceptance result.
Example task
Teaching example: an autocollimator with a polygon prism; five components are combined and expanded. Numbers are illustrative, not a formal laboratory budget.
Setup and preparation
Define items, range and target uncertainty.
Select standards and support equipment along the traceability chain.
Set up environmental monitoring, equipment files and interim checks.
Write the procedure, record template and uncertainty budget.
Example data and records
Components in arc-seconds: reference 0.05, repeatability 0.02, mounting eccentricity 0.03, environment 0.02, resolution 0.01.
Combined standard uncertainty 0.0656″.
Expanded uncertainty (k=2) 0.131″.
Calculation steps
u_c = √(0.05²+0.02²+0.03²+0.02²+0.01²) = 0.0656″; U = 2u_c = 0.131″. Correlated components need covariance terms.
Interpretation
The set-up gives about 0.131″ at k=2; a tolerance below three times that value cannot be supported by the measurement.
Verification and pitfalls
The budget must match the procedure; a new mounting needs a new budget.
The repeatability term needs enough independent repeats.
Certificate values are not automatically uncertainty components.