Home / In-house angular calibration capability
DOWNLOADABLE MEASUREMENT PLAN

In-house angular calibration capability

Build internal calibration for autocollimators, electronic levels and angle standards, including equipment, environment and uncertainty budget.

Instruments

Procedure

Calculation

The budget covers standard, repeatability, eccentricity, environment and resolution; combine in quadrature and apply the coverage factor.

Reporting

Limitations

Formal accreditation follows local metrology authority requirements; site content does not replace assessment.

STEP BY STEP

Understand the measurement step by step

  1. 01

    Define items, range and target uncertainty.

  2. 02

    Select standards and support equipment along the traceability chain.

  3. 03

    Set up environmental monitoring, equipment files and interim checks.

  4. 04

    Write the procedure, record template and uncertainty budget.

  5. 05

    Run repeatability and comparison tests to confirm capability.

u_c = √Σuᵢ²U = k · u_cPrinciple schematic · not to scale
1 / 5
  1. Step 1 · Define items, range and target uncertainty.
  2. Step 2 · Select standards and support equipment along the traceability chain.
  3. Step 3 · Set up environmental monitoring, equipment files and interim checks.
  4. Step 4 · Write the procedure, record template and uncertainty budget.
  5. Step 5 · Run repeatability and comparison tests to confirm capability.

Teaching animation: follow the stations and reference directions. Do not read dimensions or angles from the schematic.

WORKED EXAMPLE

Worked uncertainty budget for an angle standard

Teaching example with assumed data, not a customer project, instrument accuracy claim or acceptance result.

Example task

Teaching example: an autocollimator with a polygon prism; five components are combined and expanded. Numbers are illustrative, not a formal laboratory budget.

Setup and preparation

  • Define items, range and target uncertainty.

  • Select standards and support equipment along the traceability chain.

  • Set up environmental monitoring, equipment files and interim checks.

  • Write the procedure, record template and uncertainty budget.

Example data and records

  • Components in arc-seconds: reference 0.05, repeatability 0.02, mounting eccentricity 0.03, environment 0.02, resolution 0.01.

  • Combined standard uncertainty 0.0656″.

  • Expanded uncertainty (k=2) 0.131″.

Calculation steps

u_c = √(0.05²+0.02²+0.03²+0.02²+0.01²) = 0.0656″; U = 2u_c = 0.131″. Correlated components need covariance terms.

Interpretation

The set-up gives about 0.131″ at k=2; a tolerance below three times that value cannot be supported by the measurement.

Verification and pitfalls

  • The budget must match the procedure; a new mounting needs a new budget.

  • The repeatability term needs enough independent repeats.

  • Certificate values are not automatically uncertainty components.