Home / Splice flatness inspection plan for optical breadboards
DOWNLOADABLE MEASUREMENT PLAN

Splice flatness inspection plan for optical breadboards

For splice flatness and seam steps after joining multiple optical breadboards into one surface (2×1 pieces of 600×900 mm, example/teaching data). Support per the vendor's recommended array, measure the full surface on a grid, evaluate against a least-squares reference and report seam steps separately.

Instruments

Procedure

Calculation

The least-squares plane z=ax+by+c minimises squared residuals; flatness is the residual peak-to-valley. Seam step is the height difference of the nearest stations on both sides of the seam. With L=100 mm, 1″≈0.485 μm (example/teaching data).

Reporting

Limitations

Local stiffness variations of the honeycomb make high-frequency waviness between spans unmeasurable; this plan evaluates the mechanical mounting surface only, not optical figure or isolation performance.

STEP BY STEP

Understand the measurement step by step

  1. 01

    Confirm the 2×1 splice layout (example), support array and least-squares evaluation agreement.

  2. 02

    Level each board before joining; after joining, lay out a 100 mm grid over the full surface with densified stations on both sides of each seam.

  3. 03

    Read each grid line with the autocollimator to 0.1″ with a fixed sign convention.

  4. 04

    Tie all lines to a common datum into a height matrix and fit a least-squares reference plane.

  5. 05

    Evaluate splice flatness (residual peak-to-valley) and report each seam step separately.

  6. 06

    Unclamp and re-clamp the supports, then repeat key grid lines to assess clamping repeatability.

  7. 07

    Deliver the height matrix, flatness, seam step table and clamping repeatability record.

Principle schematic · not to scale
1 / 5
  1. Splice set-up · Two breadboards spliced and torqued.
  2. Aim at joint · Autocollimator aimed along the splice.
  3. Cross-joint survey · The 150 mm bridge steps across the joint.
  4. Compare readings · Joint and off-joint readings compared.
  5. Step evaluation · Height conversion gives the splice step.

Teaching animation: follow the stations and reference directions. Do not read dimensions or angles from the schematic.

WORKED EXAMPLE

Optical Breadboard Splice Flatness Inspection · Tutorial Case

Teaching example with assumed data, not a customer project, instrument accuracy claim or acceptance result.

Example task

Teaching demonstration, not a real customer project: segmented measurement of step height and alignment across a splice between two optical breadboards; data are instructional settings.

Setup and preparation

  • Splice and clamp the two breadboards on the isolation table

  • Aim the autocollimator along the splice direction

  • Survey with a 150 mm bridge crossing the joint step by step

  • Record the angular reading per segment

  • Compare heights of joint-crossing and plain segments

Example data and records

  • Bridge span L=150 mm; segment 1 autocollimator reading θ₁=+0.3″

  • Segment 2 autocollimator reading θ₂=-0.6″

  • Segment 3 autocollimator reading θ₃=+0.9″

Calculation steps

Height conversion: Δh = L·tan(θ·π/648000)×1000 (μm). Segment 1: 150×tan(+0.3″)=+0.22 μm; Segment 2: 150×tan(-0.6″)=-0.44 μm; Segment 3: 150×tan(+0.9″)=+0.65 μm. Cumulative heights: 0→+0.22→-0.22→+0.43 μm; peak-to-valley PV=+0.43−(-0.22)=+0.65 μm.

Interpretation

Tutorial result PV≈0.65 μm, demonstrating evaluation of the splice step height.

Verification and pitfalls

  • Dowels and screws torqued to specification

  • When crossing the joint the bridge feet straddle both boards

  • Do not touch the table during readings