Wafer stage guideway and flatness

Straightness, flatness and thermal stability of wafer, exposure and precision XY stages.

Instruments

Procedure

Calculation

Straightness by angle integration or direct displacement; flatness on one datum; drift as reference-point motion over time.

Reporting

Limitations

Sub-micron results depend on support, airflow and operator position; measure in the service state.

STEP BY STEP

Understand the measurement step by step

  1. 01

    Confirm cleanliness, isolation and vibration limit.

  2. 02

    Stabilise thermally and establish the datum, avoiding hand heat.

  3. 03

    Measure straightness in both planes and table flatness.

  4. 04

    Run multi-point repeatability and reversal, logging thermal drift.

  5. 05

    Report per-axis error, flatness and drift curves.

Principle schematic · not to scale
1 / 5
  1. Step 1 · Confirm cleanliness, isolation and vibration limit.
  2. Step 2 · Stabilise thermally and establish the datum, avoiding hand heat.
  3. Step 3 · Measure straightness in both planes and table flatness.
  4. Step 4 · Run multi-point repeatability and reversal, logging thermal drift.
  5. Step 5 · Report per-axis error, flatness and drift curves.

Teaching animation: follow the stations and reference directions. Do not read dimensions or angles from the schematic.

WORKED EXAMPLE

Worked 300 mm stage straightness and thermal drift example

Teaching example with assumed data, not a customer project, instrument accuracy claim or acceptance result.

Example task

Teaching example: a 300 mm travel split into five bays, angles integrated into straightness, with a reference point tracking thermal drift. Synthetic data.

Setup and preparation

  • Confirm cleanliness, isolation and vibration limit.

  • Stabilise thermally and establish the datum, avoiding hand heat.

  • Measure straightness in both planes and table flatness.

  • Run multi-point repeatability and reversal, logging thermal drift.

Example data and records

  • Bay angles in arc-seconds: +0.3, +0.2, -0.1, -0.2, -0.2 over 60 mm bays.

  • Accumulated height: 0.087, 0.145, 0.116, 0.058, 0 µm.

  • Reference point drifted 0.05 µm over 2 h, used to separate thermal effects.

Calculation steps

Δh = 0.087 µm; the accumulated range 0.145 µm is the straightness over this travel. Drift is reported separately and never subtracted from the profile.

Interpretation

Straightness is about 0.145 µm concentrated in the first bays, suggesting mounting or preload at the entry; drift is far smaller and not the cause.

Verification and pitfalls

  • Airflow and acoustic vibration disturb sub-micrometre readings; switch off non-essential equipment.

  • Let the sensor reach stage temperature; measuring on a cold start adds thermal error.

  • Keep straightness, flatness and angular error separate.