Setup: short range, short travel
Z travel is short and the measuring distance usually under 0.5 m, so a wide-field or short-range instrument is convenient. Clamp the mirror to the side of the stage with the face normal aimed at the optical axis.
Keep the microscope lamp cooling fan from blowing across the beam path; at short distances warm air is the main disturbance.
Readings: points over the travel
Start at the bottom, take a point every 3 mm and record both axes; run up and down once to check hysteresis — 11 points over 30 mm.
Teaching example: the horizontal axis drifts monotonically from 0 to +4.6″. Multiplied by a 0.5 m path this is an 11 μm-order displacement — a path amplification effect; the true lateral offset must be converted from the geometry.
Conversion and verdict
The angle times the lever arm from mirror to the measured axis gives the lateral offset of that axis — not the angle times the optical path. With a 100 mm lever arm, 4.6″ is about 2.2 μm, usually acceptable for stitching systems.
The acceptance limit is derived from the stitching specification of the instrument maker; this example uses teaching values.
Case workflow, animated
- Confirm the task and datum
- Lay out stations by span
- Acquire readings and convert
- Evaluate and verify
The animation follows the station layout and calculation order of this case. Values are teaching data.
Case data are teaching assumptions. Angle-to-displacement conversion must state the lever-arm definition; stage acceptance is defined by the instrument manufacturer documentation.