Principle
A calibrated rotation plus residual small-angle readings defines the separation of face normals.
Preparation
Define internal, external or normal-rotation angles; calibrate the angle generator and inspect face quality and beam footprint.
Workflow diagram
Procedure
- Identify faces and define the required angle.
- Align the first face as the reference.
- Rotate the reference device until the second return is within range.
- Combine reference rotation and residual readings using the defined signs.
- Repeat orientations or cycles to inspect face-form and eccentricity effects.
Calculation
Determine the signed normal rotation β. For adjacent faces of a convex prism in this geometry, A=180°−β; other setups need their own optical model.
Illustrative example
Illustration: in that geometry, a normal rotation of 90°+1″ corresponds to an internal angle of 90°−1″.
Limitations
- Normal rotation and the geometric internal angle are not always identical.
- Nonflat faces and beam-footprint changes affect the measured orientation.
Reporting
Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.