MICROCRE / OPTICAL METROLOGY400-812-7833  ·  SHANGHAI
研润 WHW
MICROCREPRECISION IN EVERY ANGLE
Home / Measuring the Deviation Angle of a Prism
MEASUREMENT METHOD

Measuring the Deviation Angle of a Prism

Measure a prism's apex angle (the included angle of its refracting faces) and its deviation angle for component inspection and refractometry. The apex angle combined with the minimum deviation yields the refractive index by the classical formula, a foundational optical metrology method. It suits high-accuracy single-prism measurement in a metrology room.

Principle

The apex angle is measured by autocollimation: with the prism on a table, the autocollimator is autocollimated on each refracting face (return image on the reticle); the reading difference gives the angle φ between face normals and A=180°−φ. The deviation angle between incident and emergent rays is measured directly by rotating the autocollimator or the table. At minimum deviation the first-order sensitivity to incidence angle vanishes, making the measurement robust. With A and the minimum deviation δmin, the refractive index is n=sin[(A+δmin)/2]/sin(A/2).

Preparation

Clean the prism and inspect the transmitting faces for chips and scratches. Clamp it on the rotary table without stress. Match the autocollimator height to the table axis; the table resolution (e.g. 1″–10″ class) should pair with an autocollimator of 0.1″–0.5″ class. Record the ambient temperature.

Workflow diagram

01Mount the prism02Autocollimate face 103Autocollimate face 204Compute apex angle05Find minimum deviation06Compute and report

Procedure

  1. Clamp the prism on the table so both refracting faces can be brought toward the autocollimator within rotation range.
  2. Rotate to autocollimate the first face (return image on the reticle) and record the table position φ1.
  3. Rotate to autocollimate the second face and record φ2; the apex angle A=180°−|φ2−φ1|.
  4. Re-measure both faces to confirm clamping stability and average the results.
  5. Observe the deviation in transmission: track the emergent beam versus table angle; the extremum is the minimum deviation δmin.
  6. Compute the refractive index if required and report A, δmin, n with conditions and uncertainty.

Calculation

Apex angle: A=180°−φ with φ=|φ2−φ1| the angle between face normals. Refractive index: n=sin[(A+δmin)/2]/sin(A/2). General deviation (teaching form): δ=i1+i2′−A with incidence i1 and emergence i2′; at minimum deviation dδ/di1=0. Conversion: 1″=π/648000 rad. The index uncertainty propagates from A and δmin via partial derivatives of the formula.

Illustrative example

Schematic · not measured data
Schematic: apex angle from two-face autocollimation and the minimum-deviation setting; values illustrative.

Teaching data, not measured: autocollimation readings φ1=30°00′10″ and φ2=90°00′15″ give φ=60°00′05″ and A=59°59′55″. With δmin=40°00′20″, the demonstration index is about n≈sin(50°00′07.5″)/sin(29°59′57.5″)≈1.5322. No real prism is implied.

Limitations

  • Clamping stress or temperature changes alter the apex angle; re-check between readings.
  • Autocollimation targets the face normal, not the edge; writing φ instead of 180°−φ inverts the geometry.
  • The deviation curve is flat near the minimum, so inadequate table resolution inflates the δmin uncertainty.

Reporting

Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.

Method reference ↗ Method reference ↗ 
STEP BY STEP

Understand the measurement step by step

Principle schematic · not to scale
1 / 5
  1. The prism on the table with the autocollimator on face 1 and the image centred on the reticle.
  2. The table indexes to face 2, φ₂ and r₂ are logged and the vertex angle appears on screen.
  3. The helium lamp lights up and the deviated spot sweeps as the table turns.
  4. The deviation curve is plotted and the extremum is flagged as δmin.
  5. The formula takes A and δmin and prints n=1.51680 digit by digit.

Teaching animation: follow the stations and reference directions. Do not read dimensions or angles from the schematic.

WORKED EXAMPLE

Vertex angle and minimum deviation of a 60° dispersing prism

Teaching example with assumed data, not a customer project, instrument accuracy claim or acceptance result.

Example task

An incoming-inspection run on a 60° dispersing prism must report the vertex-angle error, the minimum deviation and the refractive index. The prism is waxed to a precision rotary table, an autocollimator (0.01″) sights the refracting faces and the table reading gives each face normal; the minimum-deviation position is then found at the helium d line (587.6 nm).

Setup and preparation

  • Wax the prism at the centre of the table so both refracting faces can be autocollimated within the table travel, with the optical axis about 40 mm above the table surface.

  • Fix the autocollimator roughly normal to face 1 and focus until the return image is sharp and centred on the reticle.

  • Rotate the table until face 1 autocollimates, recording the table angle φ₁ and the residual r₁; repeat for face 2 to obtain φ₂ and r₂.

  • Repeat both faces once to check that the wax has not crept, then average.

  • Switch to a helium lamp with a 587.6 nm filter, rotate the table to track the deviated beam and locate the extremum of deviation against table angle.

  • Fine-step 0.02° around the extremum and lock when the reading changes by less than 0.001°, then record δmin twice.

Example data and records

  • Face 1 autocollimated: table 0.0000°, autocollimator residual +1.20″

  • Face 2 autocollimated: table 119.9992°, autocollimator residual -0.80″

  • Face 1 repeated: table 0.0002°, residual +1.00″

  • Face 2 repeated: table 119.9989°, residual -0.70″

  • Minimum deviation position (He d line, 587.6 nm): δmin reading 38.6481°

  • Two repeats: 38.6482° and 38.6479°

  • 20.1 °C; the prism is held with soft wax, no clamp on the optical faces

Calculation steps

The angle between the two face normals is φ=(φ₂−φ₁)+(r₂−r₁)=(119.9992−0.0000)+(−0.8−1.2)/3600=119.998644°. The vertex angle is A=180°−φ=60.001356°=60°00′04.88″, i.e. +4.88″ from 60°. The minimum deviation is δmin=38.6481°=38°38′53.16″. The refractive index follows from n=sin[(A+δmin)/2]/sin(A/2)=sin(49.3247°)/sin(30.0007°)=0.758416/0.500010=1.51680. Each repeated residual moved by no more than 0.3″, which is 0.3″ of vertex angle, so the average carries about 0.2″ of uncertainty. Using the nominal 60° instead of the measured vertex angle would bias the index by about 4.2×10⁻⁵, which is why the angle is measured first.

Interpretation

Vertex angle 60°00′04.88″ (error +4.88″), minimum deviation 38°38′53.16″ and n=1.51680. The vertex angle passes the ±10″ tolerance, and n agrees with the 1.5168 nominal for BK7 at 587.6 nm to better than 1×10⁻⁴, as expected for this batch.

Verification and pitfalls

  • The vertex angle combines two table readings with the autocollimator residuals, and the signs must agree: with r₂−r₁=-2.00″ the residual adds to the table rotation; subtracting it would bias the angle by 2.00″.

  • Check the table for backlash in both directions; above 2″ measure forward and backward and average, or the angle will depend on the direction of approach.

  • Never clamp the refracting faces: the soft wax is there to avoid stress birefringence and form change.

  • Confirm the extremum of minimum deviation: the reading must increase 0.5° either side; a monotonic trend means the extremum was missed.

  • The index is temperature sensitive (dn/dT about 3×10⁻⁶/K), so hold the room within 0.5 °C.