Principle
In an ideal roof prism the two half-faces split and rejoin the beam seamlessly; a roof angle deviating from 90° tilts the halves apart and produces a double image on emergence. In autocollimation the half-field reticle images are displaced across the roof-edge direction, and the angular separation is proportional to the roof-angle error. The conversion factor depends on the prism type, refractive index and incidence angle on the roof faces, and must follow the design data or manual for that prism rather than a generic coefficient.
Preparation
Identify the prism type (right-angle roof, Pechan roof, etc.) and the source of its design conversion factor. Clean all transmitting faces and the roof edge, verifying the edge is unchipped. The autocollimator aperture must cover both half-apertures with uniform return brightness. Support the prism stably and avoid vignetting either half.
Workflow diagram
Procedure
- Consult the prism datasheet for the relation between double-image separation and roof-angle error.
- Clean and mount the prism stably, noting the roof-edge orientation.
- Align the autocollimator and focus the reflected image; confirm both half-fields are fully visible.
- Read the separation τ between the half-images perpendicular to the roof edge, in arcsec with sign.
- Convert τ to the roof-angle error ε with the design factor and determine whether the angle exceeds or falls short of 90°.
- Repeat at a different incidence to validate the conversion and report ε, the criterion and uncertainty.
Calculation
General form (example, verify per type): ε=τ/K, where K is the double-image magnification factor of that prism, varying with type and index (for some right-angle roof prisms on the order of 4n·cosβ, β being the incidence on the roof faces). The report must cite the source of K. If only a range for K is documented, evaluate with the least favourable value and include it in the uncertainty.
Illustrative example
Teaching data, not measured: a datasheet quotes K=6.0 (example). If the autocollimator reads τ=+12″, the roof-angle error is ε=+2.0″ as a demonstration calculation, the sense defined by the datasheet. No real prism or instrument accuracy is implied.
Limitations
- Applying a conversion factor from a different prism type invalidates the result.
- Vignetting one half-aperture hides the double image and falsely suggests zero error.
- Scatter from a chipped or contaminated roof edge misread as angular error.
Reporting