The same instrument can produce results differing by micrometres depending on the setup. Reflector, base span, cosine error and Abbe error are the four factors that must be made explicit.

cos errormotion axismeasuring line
Δh = L · tan θ

Animated explanation

LEDBeamsplitterImage planeReturn beam retraces and splits againSchematic · not to scale · not measured data
1 / 4
  1. Establish the geometry
  2. See how one reading forms
  3. Change a condition and observe
  4. Return to the formula and its limits

The animation explains relationships and directions. Read numbers from the formula, the standard and your records.

The reflector is part of the measuring chain

The autocollimator reads the mirror normal direction. Surface form error, coating unevenness and mounting stress alter the local normal and enter the reading directly.

Precision work should use a mirror with a verified surface and state its grade in the report. At long distance, check that the mirror size covers the spot.

Base span and cosine error

Converting angle to height needs the base span L: Δh = L·tanθ. Here L is the distance between the two supports of the base, not the optical path and not the part length.

If the measuring line is not parallel to the motion axis, readings shrink by cos of the misalignment. At 1° the effect is about 0.015%; at 5° about 0.4%, which accumulates over long travel.

Abbe error and offset

When the measuring line is offset from the geometric feature under test, any angular motion is amplified into extra displacement. This is the Abbe error.

Lay out the path so that the measuring line passes through or near the feature. If an offset is unavoidable, record the offset distance and explain it when interpreting results.

Common accessories and what they do

A right-angle adapter or pentaprism folds the path by 90° for squareness between a guideway and a column. A polygon provides known angular increments for indexing and angle references.

Magnetic bases, adjustable mounts and bridge bases determine how repeatable a setup is. Accessory errors are not covered by the instrument specification and must be checked separately.