Home / Encoder positioning accuracy: polygon and autocollimator comparison
CASE 13 · TEACHING DATA

Encoder positioning accuracy: polygon and autocollimator comparison

High-grade encoder positioning error must be compared against a mechanical angle standard. Mount a 23-face polygon on the encoder shaft, read face angles with an autocollimator, and compare point by point with the encoder output to obtain the full-revolution error curve.

ObjectAbsolute angle encoder, rated system accuracy ±2″
Reference23-face polygon (grade 0) + 0.1″ class autocollimator
Points23 points per revolution, 2 runs each direction
Environment20 ± 0.5 °C, isolated table
repeatspanmirrorangleu_c = sqrt(sum u_i^2)

Setup: coaxiality first

Mount the polygon on the encoder shaft via an arbor and centre eccentricity within 5 μm; the autocollimator stands on an independent support and does not rotate. Eccentricity adds a second harmonic to the result and must be removed first.

Aim the optical axis at face-centre height and warm up the instrument for at least 30 minutes before sampling.

Readings: face-by-face comparison

Stop at each face, record the encoder output and the autocollimator reading; after 23 points repeat in reverse. The forward/reverse difference shows backlash and friction; the average feeds the positioning error evaluation.

Teaching example: at face 7 the encoder reads 109°33′54.8″ and the autocollimator converts to 109°33′53.1″, a deviation of +1.7″ at that point.

Calculation: closure and the error curve

Polygon face angles have calibration certificate values; correct the autocollimator readings with them and take point-wise differences to obtain the encoder error curve. The closure error of a full revolution should stay below twice the measurement repeatability, otherwise recheck clamping and temperature.

Fourier decomposition of the curve: the first harmonic usually comes from residual eccentricity, higher orders reflect the encoder grating error itself.

Limits of the conclusion

The method evaluates the positioning error of the shaft-plus-encoder assembly; isolating the encoder alone needs a higher-grade spindle or a multi-head separation method.

Example readings and curves are teaching assumptions that illustrate the calculation and are not measured results for any specific encoder product.

Case workflow, animated

ABCDΔ₄₁Closure c = ΣΔ should approach 0; c/4 is a conditional correctionSchematic · not to scale · not measured data
1 / 4
  1. Confirm the task and datum
  2. Lay out stations by span
  3. Acquire readings and convert
  4. Evaluate and verify

The animation follows the station layout and calculation order of this case. Values are teaching data.

ConclusionAfter the 23-point comparison the peak-to-peak positioning error is 3.8″ with a 1.2″ first harmonic; after removing residual eccentricity the encoder body error is about ±1.5″, within the ±2″ rating. (Teaching example data)

Case data are teaching assumptions. Encoder acceptance follows the manufacturer specification and the test conditions agreed by both parties; polygon grade, autocollimator accuracy and environment together set the achievable uncertainty.