Home / 200 mm precision stage: separating accumulation from reversal jump
CASE 07 · TEACHING DATA

200 mm precision stage: separating accumulation from reversal jump

Unidirectional pitch accumulation is often modest; what hurts bidirectional positioning is the jump at reversal. Separate the two sampling modes to tell geometry from preload.

Object200 mm stage, 5 segments at 40 mm, unidirectional and bidirectional sets
InstrumentDual-axis autocollimator, 0.05″ resolution, 40 mm mirror base, tracking mount
Conditions20±0.3 °C, air supply stable at 0.4 MPa, 30 min warm-up
Criterionpitch ≤2″/200 mm, bidirectional difference ≤1″
DUAL AXIS+Y-Y+X field-X field

Why two sampling modes

Unidirectional sampling maps geometry: read every 40 mm, accumulate into height, 2.46 μm over five segments — normal for a 200 mm stroke.

Bidirectional sampling exposes something else: approaching one position from either side gives readings differing by 0.8″. That is backlash or weak preload, not profile error.

Reading the reversal jump

At each point, drive forward 40 mm, settle and read; continue forward 40 mm, then reverse back and read again. No other motion in between.

A repeatable, stable jump points to mechanical clearance; a drifting one suggests air-pressure fluctuation or encoder signal quality.

Conclusion and order of fixes

Fix the reversal jump before touching the profile: the reversal error enters bidirectional accuracy directly, while 2.46 μm of accumulation is acceptable for many uses.

Re-measure after adjusting preload. If the jump drops below 0.3″ with accumulation unchanged, no geometric compensation is needed; only then consider a compensation table.

Case workflow, animated

ABCDΔ₄₁Closure c = ΣΔ should approach 0; c/4 is a conditional correctionSchematic · not to scale · not measured data
1 / 4
  1. Confirm the task and datum
  2. Lay out stations by span
  3. Acquire readings and convert
  4. Evaluate and verify

The animation follows the station layout and calculation order of this case. Values are teaching data.

ConclusionUnidirectional accumulation of 2.46 μm (about 0.4″) is acceptable; the 0.8″ bidirectional difference uses most of the 1″ budget, so adjust preload and re-measure before any software compensation.

Teaching data. Real stages also involve encoder error, air-bearing stiffness and load variation; this case covers angles only and is not a final positioning verdict.

Recommended instruments