Principle
A plane mirror is attached to the spindle face, roughly normal to the axis, while a two-axis autocollimator reads its tilt in two orthogonal directions. During rotation the tilt sequence contains a once-per-revolution component corresponding to spindle-axis tilt plus the contribution of mirror mounting eccentricity. The distinction matters: pure translation of a plane mirror does not change the reading, only tilt does; mounting eccentricity traces a cone of the mirror normal, also appearing as a once-per-revolution harmonic. The classical remedy is measuring over full turns in both directions or at two positions to separate eccentricity from genuine tilt error motion.
Preparation
Clean the spindle face and prepare a rigid mirror mount with minimal radial and axial runout after installation. Set the autocollimator at a suitable distance with return light throughout the rotation. Choose sampling: continuous or stepwise at discrete angles. Record speed, temperature and support conditions such as bearing preload changes.
Workflow diagram
Procedure
- Mount the mirror on the spindle face, aligning its initial tilt near mid-range of the autocollimator.
- Aim at the mirror centre and confirm both readings stay in range through the full rotation.
- Rotate the spindle at the set speed or step and record angle plus both tilt readings synchronously.
- Acquire at least one full revolution, usually averaging several to suppress random noise.
- Perform harmonic analysis on both sequences: extract once- and higher-order harmonics and use two-position or two-direction data to separate mounting eccentricity from true tilt error.
- Report polar curves of tilt error versus angle with peak indices and measurement conditions.
Calculation
The two readings θx(φ) and θy(φ) combine as θ(φ)=√(θx²+θy²). First-harmonic fits θx=a1·cosφ+b1·sinφ+c1 (likewise θy) describe the composite of spindle tilt and mirror eccentricity; with two-position measurements half the difference of the two first harmonics yields the pure tilt component (classical treatment, derived per layout). Peak indices take the maximum of θ(φ) or the maximum deviation from the mean axis.
Illustrative example
Teaching data, not measured: suppose mirror eccentricity contributes a first harmonic of 2″ and true spindle tilt 0.8″. A single-position measurement reads their vector composite (a demonstration magnitude of about 2.6″); two orthogonal positions separate the true tilt of about 0.8″. The numbers only illustrate the separation.
Limitations
- Attributing the whole first harmonic to spindle tilt while ignoring mirror eccentricity.
- Incomplete or uneven sampling over the revolution distorts the harmonic analysis.
- A compliant mirror mount mixes structural vibration into the readings.
Reporting
Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.