MEASUREMENT METHOD

Assessing Vibration Effects on Arcsecond Readings

Quantify how ambient vibration affects arcsecond-class autocollimator readings and judge whether the environment meets the intended resolution and repeatability. It is used for site acceptance, diagnosing anomalous readings and selecting measurement windows, and informs isolation measures or reading strategies such as time averaging.

Principle

In a truly quiet environment arcsecond readings are limited only by instrument noise; ambient vibration imposes angular motion on the reflector or the instrument and the readings scatter about the true value, visible as image jitter. With reflector and instrument stationary, continuous sampling at fixed intervals yields a time series whose standard deviation reflects combined vibration and noise; spectral analysis identifies dominant frequencies (mains 50 Hz, rotating machinery) and separates vibration from random noise and drift. Comparing series from different periods, such as machinery on and off, quantifies a specific source's contribution.

Preparation

Set reflector and autocollimator in working condition with a static target, readings near mid-range. Choose interval and duration (e.g. 1 s sampling for 5–10 min) covering several cycles of the main sources. Log the state of nearby equipment during the window (on/off list). For spectra, sample at several times the highest frequency of interest.

Workflow diagram

01Static target02Sample time series03Statistics and drift04Spectrum identifies sources05Assessment and advice06Compare windows and conclude

Procedure

  1. Position reflector and instrument, confirming stable return and mid-range readings.
  2. Record readings at a fixed interval over the chosen window while logging equipment state.
  3. Repeat in a control window, e.g. with machinery off, where possible.
  4. Compute the mean, standard deviation and drift trend (such as first-versus-last-segment means) of each series.
  5. Perform spectral analysis to identify dominant frequencies and match them to candidate sources.
  6. Compare standard deviations between windows and report the vibration contribution, conformity verdict and recommendations on isolation or reading strategy.

Calculation

For the series {ri}, i=1..N: mean r̄=Σri/N and standard deviation s=√[Σ(ri−r̄)²/(N−1)]. Drift is expressed as the difference of first- and last-segment means. Spectral amplitudes A(f) come from the discrete Fourier transform. The vibration contribution is the standard-deviation difference between windows: Δv=s_on−s_off (example). If s exceeds a set multiple of the target resolution, e.g. three times, the environment fails and measures are required.

Illustrative example

Schematic · not measured data
Schematic: time series on a static target and spectral components; values illustrative.

Teaching data, not measured: a machinery-on window gives a standard deviation of 0.08″ and an off window 0.02″, so the vibration contribution is about 0.06″ (demonstration). With a target resolution of 0.01″ and an agreed threshold of three times (0.03″), the window fails for measurement. An example spectrum shows a 25 Hz component matching a fan speed (illustrative scenario).

Limitations

  • Mistaking vibration-induced scatter for genuine angular change of the measurand.
  • Too coarse an interval or too short a window misses low-frequency and intermittent sources.
  • Mixing results from vibrating and quiet windows in one report without labels.

Reporting

Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.

Method reference ↗ Method reference ↗ 
STEP BY STEP

Understand the measurement step by step

Principle schematic · not to scale
1 / 6
  1. The mirror and autocollimator static mid-guideway while the reading scrolls on screen.
  2. With machines stopped the trace is almost a straight line with fine noise.
  3. The crane passes, the trace shakes violently and the 1.80″ range is marked.
  4. The press starts and the swing grows to a 4.60″ range.
  5. The statistics card shows 0.03″, 0.42″, 1.20″ and the 0.13″ averaged uncertainty.
  6. A verdict label lights up: pause measurement in this window.

Teaching animation: follow the stations and reference directions. Do not read dimensions or angles from the schematic.

WORKED EXAMPLE

Effect of shop vibration on autocollimator arcsecond readings

Teaching example with assumed data, not a customer project, instrument accuracy claim or acceptance result.

Example task

During a 200 mm-span guideway check the autocollimator readings jitter whenever the travelling crane or the press runs. To quantify the effect, the mirror and instrument are left static and 10 consecutive samples are taken with the machines stopped, with the crane passing and with the press running; the standard deviations and single-station repeatability then decide whether to wait or to average more.

Setup and preparation

  • Set up the mirror and autocollimator in the normal measuring position, static, with a stable beam and the reading mid-range.

  • With the machines stopped, log 10 consecutive readings as the instrument noise and floor background, noting the state of nearby equipment.

  • Repeat 10 samples while the crane passes, and make three repeats at one station in the same period to record the range.

  • Take another 10 samples with the press running to complete three comparison sets.

  • Compute mean, standard deviation and range for each set and compare the scatter caused by vibration.

  • Record 21.0 °C; the station is mid-guideway near the aisle and has no isolation pad.

Example data and records

  • Machine stopped: 10 consecutive readings, mean +0.02″, standard deviation 0.03″

  • Travelling crane passing: mean +0.05″, standard deviation 0.42″, range 1.80″

  • Three repeats at one station in that period: +0.68″, −1.12″, +0.31″, range 1.80″

  • Press running: mean −0.18″, standard deviation 1.20″, range 4.60″

  • Mirror and autocollimator both static at 200 mm, no isolation pad

  • 21.0 °C; the vibration station is mid-guideway, close to the shop aisle

Calculation steps

Static noise σ₀=0.03″, crane σ=0.42″ which is 14 times higher, press σ=1.20″ which is 40 times higher. Taking one standard deviation as the single-reading uncertainty gives 0.03″, 0.42″ and 1.20″. Averaging N readings reduces the random part to σ/√N: 10 averages during crane passage give 0.42/√10=0.13″ and 20 averages give 0.09″. Converted over the 200 mm span the height uncertainty is 200×tan(0.13″)×1000=0.13 μm. However the three-repeat range of 1.80″ is far beyond the 0.50″ allowance, so the disturbance is not purely random and averaging cannot remove a low-frequency sway. The conclusion is therefore to stop measuring while the crane passes and re-measure afterwards rather than to average through it.

Interpretation

Static background 0.03″, crane passage 0.42″ (14 times higher) and press running 1.20″. Ten averages reduce the random part to 0.13″, but the range of 1.80″ exceeds the allowance, so crane and press periods are declared unmeasurable and measurement waits for shutdown.

Verification and pitfalls

  • Standard deviation must be read together with the range: σ=0.42″ looks averageable, but a 1.80″ range reveals low-frequency sway that averaging cannot fix.

  • Record the sampling interval: 0.5 s and 5 s intervals differ by about 30 % in standard deviation, so data from different intervals cannot be compared.

  • Vibration is directional: only angular vibration about the mirror normal enters the reading, so linear vibration levels must not be used as an estimate.

  • An isolation pad helps at high frequency but barely affects the 1–3 Hz sway from a crane, so prefer a shutdown window.

  • Take all three sets on the same day and station; comparing across days mixes in temperature and support changes.