Home / Indexing table, both directions: how much of 3.5″ is backlash
CASE 20 · TEACHING DATA

Indexing table, both directions: how much of 3.5″ is backlash

A unidirectional test folds backlash into indexing accuracy. Measuring both directions shows whether to compensate or repair.

Object12-station NC table, 90:1 ratio
Reference24-face polygon, grade 1
Instrument0.1″ autocollimator
Result3.5″ one-way, 1.8″ max reversal
DUAL AXIS+Y-Y+X field-X field

Measurement plan

Polygon centred within 5 μm; three repeats each direction per face.

Run a full-circle closure first.

Separating backlash

A 1.8″ difference against 0.4″ repeatability indicates mechanical backlash.

After removing it, indexing spread falls to 2.1″ and only that part suits compensation.

Conclusion

Adjust the worm gear mechanically, then compensate; a table alone leaves the reversal error.

Case workflow, animated

AutocollimatorNominal facet angles ± deviationFull-circle closure: Σ deviations → 0Schematic · not to scale · not measured data
1 / 4
  1. Confirm the task and datum
  2. Lay out stations by span
  3. Acquire readings and convert
  4. Evaluate and verify

The animation follows the station layout and calculation order of this case. Values are teaching data.

ConclusionBacklash 1.8″; residual indexing spread 2.1″. Repair first, compensate second.

Teaching example. Follow the maker manual for disassembly and preload.