12-position rotary table: 3.5″ peak-to-valley and a reversal table

Indexing error is not random noise; it repeats with angular position. A 12-position deviation curve both judges acceptance and generates the compensation table.

ObjectNC indexing table, every 30° over 360°, both directions
Instrument0.1″ autocollimator plus a calibrated 12-face polygon on the table centre
Conditions20±0.5 °C, 30 min warm running to settle the lubricant film
Criterion±2″ anywhere, ≤5″ cumulative
DUAL AXIS+Y-Y+X field-X field

Why the polygon must be calibrated

The polygon has its own error. Treating it as absolute gives table error plus polygon error, inseparable. A calibrated polygon lets you subtract the known part.

Without a certificate, reverse the polygon 180° and average the two runs to cancel its first harmonic. That handles part of the error only, not a substitute for calibration.

What was measured

The twelve deviations are −1.2, +0.6, +1.8, +0.4, −0.9, −1.5, −0.3, +1.1, +2.0, +0.5, −1.0, −1.5 arcseconds: 3.5″ peak-to-valley, near-zero mean.

The curve undulates slowly without alternating jumps, so the worm mesh is continuous and the error is accumulated helix error rather than one bad tooth.

Compensation and verification

Negating the deviations into the control brings residuals under 1.8″. The table only holds at measured positions; intermediate points are interpolated and not necessarily linear.

After compensation, compare directions again: if the reversal difference grows from 0.4″ to 0.9″, the table has absorbed backlash and the reverse segment needs re-identification.

Case workflow, animated

AutocollimatorNominal facet angles ± deviationFull-circle closure: Σ deviations → 0Schematic · not to scale · not measured data
1 / 4
  1. Confirm the task and datum
  2. Lay out stations by span
  3. Acquire readings and convert
  4. Evaluate and verify

The animation follows the station layout and calculation order of this case. Values are teaching data.

Conclusion3.5″ peak-to-valley with +2.0/−1.5″ exceeds the ±2″ limit. A negated compensation table brings residuals to 1.8″, but re-check the reverse segment so backlash is not compensated away.

Teaching data. Real verification follows JJF or ISO 230-1 with multiple points, directions and repeats; a single 12-position polygon run cannot conclude a full circle.