JB/T 8232-2024 · Autocollimators
Product-standard candidate; confirm model applicability and inspection records. No conformity claim.
Official directory ↗Wide-field electronic autocollimators

Select using tolerance, distance, references and environmental conditions.
PDF for reading, editable Word; generated from the same source data.
The wide-field electronic autocollimator series is intended for laboratory applications requiring a large angular measuring range. It supports angular measurement of large equipment and complex structures in scientific research and industrial development.
Machine tools: guideway straightness, parallelism and perpendicularity, and flatness of large surface plates. Optics: assembly, alignment and prism-angle measurement. Metrology and research: verification and calibration of measuring instruments and tools, precision angle measurement, dynamic monitoring of angular position, and angular or straightness-error measurement of moving products. Marble inspection: straightness, parallelism, perpendicularity and flatness. Encoders: angular-encoder positioning accuracy and repeatability.
| Objective focal length | 100mm |
| Clear aperture | 38mm |
| Maximum measuring distance | 2m |
| Field of view (X, Y) | 6000*4500″ |
| Resolution | 0.1″, 0.01″, 0.001″, 0.0001″ adjustable (arc sec. arcseconds ) |
| Indication accuracy | ±1000″≤ ±2″; Full range≤ ±3″ |
| Repeatability | ≤ 0.5″ |
| Centre wavelength | 625nm |
| Measurement frequency response | ≤ 30HZ(Hz TYP.) |
| Light source type | Semiconductor LED light source |
| Communication | USB2.0 |
| Third-party communication interface | RS-232 |
| Display method | Software display on a standard PC |
| Power supply | Integrated power supply |
| Instrument dimensions (L * W * H) | 157*115*105mm |
| Main unit weight | 1.81kg |
Original source ↗ · 2026-11-14
Resolution is not accuracy. Preserve range conditions and confirm missing data with the engineering team.
Four steps: where the light comes from, where it returns, how the reading appears.
The animation shows the optical path and how a reading arises. It does not represent the resolution, range or accuracy of this model.
Wing-box and fuselage jigs spanning 3–15 m. Each locator carries a mirror; one optical axis is established from the jig datum and pitch/yaw read point by point to hold 0.05 mm assembly deviation.
Optical Alignment↗After the mirror sits in its cell, watch the returned image and adjust pitch and yaw screws into 5–30″; lock and re-check drift.
Optical Alignment↗Rotate the group and watch the image wobble; the radius of the traced circle gives decentre. Judge element by element so error does not accumulate in the assembly.
Optical Alignment↗References to check for selection, specifications and methods. Listing does not certify this model or prove testing to a standard; confirm the edition and applicable clauses in the technical agreement.
Product-standard candidate; confirm model applicability and inspection records. No conformity claim.
Official directory ↗Application, geometrical specification or evaluation reference; not a claim of instrument conformity.
Official directory ↗Application, geometrical specification or evaluation reference; not a claim of instrument conformity.
Official directory ↗Application, geometrical specification or evaluation reference; not a claim of instrument conformity.
Official directory ↗Application, geometrical specification or evaluation reference; not a claim of instrument conformity.
Official directory ↗Instrument-selection technical report, not a manufacturing or product-certification standard.
Official directory ↗Application, geometrical specification or evaluation reference; not a claim of instrument conformity.
Official directory ↗Application, geometrical specification or evaluation reference; not a claim of instrument conformity.
Official directory ↗Application, geometrical specification or evaluation reference; not a claim of instrument conformity.
Official directory ↗GB/ISO entries mainly reference application methods. JB/T 8232-2024 is a product-standard candidate; model applicability and inspection records require confirmation. No conformity claim is made.