Home / Calibrating Index Errors with a Polygon
MEASUREMENT METHOD

Calibrating Index Errors with a Polygon

Determine the index errors of digital rotary tables, dividing heads or optical index tables and derive corrections for each working position. It serves periodic verification and acceptance of indexing equipment. The polygon provides a material circular standard with closure, enabling arcsecond-level calibration with an autocollimator.

Principle

The normals of a regular polygon's faces are nominally spaced at 360°/n. The polygon is mounted coaxially on the table under test while the autocollimator stays fixed on one face. After each nominal rotation of 360°/n the next face arrives and the autocollimator reads its angular change from the optical axis, the table's index deviation at that position. The polygon's own face errors superpose on the table error; the closure condition (deviations summing to zero) separates them, and rigorous procedures use cross-calibration permutations or corrections from the polygon's certificate.

Preparation

Confirm the polygon face count (commonly 12 or 24) matches the table's nominal indexing and check the polygon certificate. Mount the polygon coaxially within the specified tolerance and clean the faces. Set the autocollimator at the face centre height and check return light over the full circle. Ensure stable temperature and no draughts.

Workflow diagram

01Mount polygon coaxially02Set up autocollimator03Step and read faces04Closure check05Compute index errors06Corrections and uncertainty

Procedure

  1. Mount and align the polygon, then rotate the table to check return light and reading range on every face.
  2. Zero on face 1 and record the initial autocollimator reading.
  3. Step the table by the nominal index angle, recording the table indication and autocollimator reading at each face.
  4. Complete the circle and return to face 1 to check the closure error.
  5. Separate table errors from polygon face errors using the closure condition, or apply certificate corrections.
  6. Report index errors, corrections and uncertainty with the calibration conditions.

Calculation

Per-position deviation e_i=A_i−A_1 (autocollimator reading difference with sign). Closure requires Σe_i=0; a residual indicates a faulty reading. Ignoring polygon errors, the table index error is e_i; rigorously e_i=t_i+p_i with table error t_i and polygon adjacent-face error p_i, separated by permutation comparison or certificate data. Sign conventions must be unified before reporting.

Illustrative example

Schematic · not measured data
Schematic: polygon steps with the table, autocollimator fixed; values illustrative.

Teaching data, not measured: calibrating a table with a 12-face polygon, the first three reading differences are +0.8″, −1.2″ and +0.4″. Neglecting polygon errors and assuming closure, these are the index errors at those positions, and the full-circle deviations must sum to zero. The numbers only illustrate the arithmetic and closure check.

Limitations

  • Polygon eccentricity drifts the return image across faces and mixes mounting error into readings.
  • Ignoring the polygon's own face errors can be commensurate with the error under test.
  • A sign inversion applies corrections the wrong way and voids the whole calibration.

Reporting

Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.

Method reference ↗ Method reference ↗ 
STEP BY STEP

Understand the measurement step by step

Principle schematic · not to scale
1 / 6
  1. The polygon sits on the table and the autocollimator faces face 1 with the image centred.
  2. The table indexes in 30° steps and the readings drop into a table one by one.
  3. After a full circle the first face returns and the reading closes at zero.
  4. The polygon is turned 180° on the table and the second run adds with the opposite sign.
  5. Subtracting the runs yields the table curve; adding them yields the polygon curve.
  6. Both curves are labelled 0.58″ and 1.58″ and the verdict card appears.

Teaching animation: follow the stations and reference directions. Do not read dimensions or angles from the schematic.

WORKED EXAMPLE

Calibrating a precision rotary table with a 12-face polygon

Teaching example with assumed data, not a customer project, instrument accuracy claim or acceptance result.

Example task

A rotary table with a ±1″ tolerance is calibrated after overhaul using a 12-face polygon. The polygon is mounted coaxially on the table centre and an autocollimator (0.01″) is fixed on one face while the table indexes 30° at a time. To separate the polygon face-angle error from the table error, one full circle is measured from face 1 and a second from face 12 (180° away); subtracting the runs cancels the even harmonics of the polygon, adding them yields the polygon error.

Setup and preparation

  • Mount the polygon on the spindle at the table centre and centre it with a dial gauge, keeping radial run-out within 3 μm.

  • Set the autocollimator beside the table, aim at the centre of face 1, focus for a sharp image and place the reading mid-range.

  • Index the table forward in 30° steps, wait 3 s per step and record both the table display and the autocollimator reading at 12 positions.

  • Return to face 1 and re-read to confirm a closure of 0.00″; investigate any looseness before continuing if it is not zero.

  • Rotate the polygon by 180° relative to the table (starting from face 7) and repeat a full circle of 12 positions.

  • Record 20.2 °C and avoid touching the polygon or the table throughout.

Example data and records

  • Position 0° (face 1): first run +0.00″, second run from 180° +0.00″

  • Position 30° (face 2): first run +0.48″, second run from 180° +0.38″

  • Position 60° (face 3): first run +0.86″, second run from 180° +0.62″

  • Position 90° (face 4): first run +1.02″, second run from 180° +0.58″

  • Position 120° (face 5): first run +0.66″, second run from 180° +0.18″

  • Position 150° (face 6): first run +0.24″, second run from 180° -0.16″

  • Position 180° (face 7): first run -0.18″, second run from 180° -0.46″

  • Position 210° (face 8): first run -0.55″, second run from 180° -0.59″

  • Position 240° (face 9): first run -0.94″, second run from 180° -0.62″

  • Position 270° (face 10): first run -1.08″, second run from 180° -0.40″

  • Position 300° (face 11): first run -0.68″, second run from 180° -0.06″

  • Position 330° (face 12): first run -0.26″, second run from 180° +0.10″

  • Closure: after 12 steps the first face is aligned again and reads 0.00″, so the circle closes

  • 20.2 °C; polygon certificate ±1.6″ face-angle error, rotary table tolerance ±1″

Calculation steps

The 12 readings of the first run average -0.04″, the mounting zero offset; after removal the residuals are +0.04 → +0.52 → +0.90 → +1.06 → +0.70 → +0.28 → -0.14 → -0.51 → -0.90 → -1.04 → -0.64 → -0.22 ″ and the combined single-run error is PV=2.10″. The second run averages -0.04″, giving residuals +0.04 → +0.42 → +0.66 → +0.62 → +0.22 → -0.12 → -0.42 → -0.55 → -0.58 → -0.36 → -0.02 → +0.14 ″. Because the second run starts 180° away, the table error changes sign while the polygon error does not, so table error Tᵢ=(r1ᵢ−r2ᵢ)/2=+0.00 → +0.05 → +0.12 → +0.22 → +0.24 → +0.20 → +0.14 → +0.02 → -0.16 → -0.34 → -0.31 → -0.18 ″ with PV=0.58″ and polygon error Pᵢ=(r1ᵢ+r2ᵢ)/2=+0.04 → +0.47 → +0.78 → +0.84 → +0.46 → +0.08 → -0.28 → -0.53 → -0.74 → -0.70 → -0.33 → -0.04 ″ with PV=1.58″. The two combine by harmonics, and the single-run value 2.10″ agrees with a √(T²+P²) estimate, so the separation is self-consistent. The polygon error 1.58″ matches its ±1.6″ certificate, so the artefact is in good condition.

Interpretation

Rotary table indexing error PV=0.58″, inside the ±1″ tolerance, so the table passes. Polygon face-angle error PV=1.58″, matching the ±1.6″ certificate, so it remains usable as a reference.

Verification and pitfalls

  • The circle must close: after 12 steps face 1 reads the starting value, 0.00″ here; a closure above 0.1″ means the polygon moved during the run.

  • The separation must be self-consistent: table 0.58″ and polygon 1.58″ cannot both approach the single-run 2.10″, which would mean the runs did not really start 180° apart.

  • Centre the polygon within 3 μm; eccentricity produces a once-per-revolution harmonic that is easily mistaken for table error.

  • Wait for the table to settle at every step: servo creep before lock shifts a single reading by more than 0.1″.

  • Keep the polygon and table at the same temperature; holding the polygon for 30 s already causes drift of about 0.2″.