Principle
In a truly quiet environment arcsecond readings are limited only by instrument noise; ambient vibration imposes angular motion on the reflector or the instrument and the readings scatter about the true value, visible as image jitter. With reflector and instrument stationary, continuous sampling at fixed intervals yields a time series whose standard deviation reflects combined vibration and noise; spectral analysis identifies dominant frequencies (mains 50 Hz, rotating machinery) and separates vibration from random noise and drift. Comparing series from different periods, such as machinery on and off, quantifies a specific source's contribution.
Preparation
Set reflector and autocollimator in working condition with a static target, readings near mid-range. Choose interval and duration (e.g. 1 s sampling for 5–10 min) covering several cycles of the main sources. Log the state of nearby equipment during the window (on/off list). For spectra, sample at several times the highest frequency of interest.
Workflow diagram
Procedure
- Position reflector and instrument, confirming stable return and mid-range readings.
- Record readings at a fixed interval over the chosen window while logging equipment state.
- Repeat in a control window, e.g. with machinery off, where possible.
- Compute the mean, standard deviation and drift trend (such as first-versus-last-segment means) of each series.
- Perform spectral analysis to identify dominant frequencies and match them to candidate sources.
- Compare standard deviations between windows and report the vibration contribution, conformity verdict and recommendations on isolation or reading strategy.
Calculation
For the series {ri}, i=1..N: mean r̄=Σri/N and standard deviation s=√[Σ(ri−r̄)²/(N−1)]. Drift is expressed as the difference of first- and last-segment means. Spectral amplitudes A(f) come from the discrete Fourier transform. The vibration contribution is the standard-deviation difference between windows: Δv=s_on−s_off (example). If s exceeds a set multiple of the target resolution, e.g. three times, the environment fails and measures are required.
Illustrative example
Teaching data, not measured: a machinery-on window gives a standard deviation of 0.08″ and an off window 0.02″, so the vibration contribution is about 0.06″ (demonstration). With a target resolution of 0.01″ and an agreed threshold of three times (0.03″), the window fails for measurement. An example spectrum shows a 25 Hz component matching a fan speed (illustrative scenario).
Limitations
- Mistaking vibration-induced scatter for genuine angular change of the measurand.
- Too coarse an interval or too short a window misses low-frequency and intermittent sources.
- Mixing results from vibrating and quiet windows in one report without labels.
Reporting
Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.