Home / Indication error, linearity and repeatability are three different numbers
PRINCIPLE NOTE

Indication error, linearity and repeatability are three different numbers

A catalogue line reading "accuracy ±1″" leaves three questions open: over which angular range, against which reference, and with what scatter. Only separate figures can be compared.

DUAL AXIS+Y-Y+X field-X field
e = θind − θref; linearity = max|e|; repeatability = max − min at one point

Animated explanation

Ideal line: angle scales linearlyDeviation = indication errorLinearity, repeatability, resolution assessed separatelyTarget angleSchematic · not to scale · not measured data
1 / 4
  1. Establish the geometry
  2. See how one reading forms
  3. Change a condition and observe
  4. Return to the formula and its limits

The animation explains relationships and directions. Read numbers from the formula, the standard and your records.

Indication error is signed deviation

Compare the reading with the reference angle from a polygon, an interferometric angle base or a calibrated table; the difference is the indication error and must keep its sign.

Only the shape of that error curve tells you whether a specific working point is usable. A single ±1″ collapses the whole curve into one number.

Linearity is the shape of that curve

Uniform sampling across the range shows whether the error grows proportionally (scale or focal length) or is worst at the ends (uncorrected sine error).

Always state range and number of points, or vendor figures cannot be compared.

Repeatability tells you whether one reading is enough

Read n times at the same position without re-clamping and report the range or standard deviation. Poor repeatability makes linearity irrelevant; check source, airflow, bracket and stability criterion first.

A working rule: repeatability no worse than one fifth of the tolerance, otherwise measurement dominates the uncertainty.