Setup: map the two axes to the two planes
The vertical axis of a dual-axis autocollimator reads angular change in the vertical plane, the horizontal axis in the horizontal plane. Seat the mirror carrier flat on the guideway contact surface with side location so the span direction follows the guideway.
Place the instrument on a stable table along the guideway extension, axis at mirror-centre height, avoiding large pitch angles that add cosine error.
Readings and reconstruction: segment integration
Record both axes at each span; integrate the vertical sequence into the vertical profile and the horizontal sequence into the horizontal profile. Δh = L·tan(θ): at a 50 mm span, 1″ equals about 0.24 μm of height difference.
Teaching example: eight vertical-axis readings (″): 0.0, +0.4, +0.7, +0.3, -0.2, -0.5, -0.3, -0.1; the integrated profile shows about 0.38 μm peak-to-valley.
Evaluation: minimum zone, not end-point line
Semiconductor practice evaluates by minimum zone: two parallel lines bracket the profile and the smallest separation is the straightness. End-point evaluation reads high, least-squares midline reads low; the report must state the algorithm used.
Conclude per plane: in this example 0.38 μm vertical and 0.29 μm horizontal (teaching values), both within 0.5 μm.
Case workflow, animated
- Confirm the task and datum
- Lay out stations by span
- Acquire readings and convert
- Evaluate and verify
The animation follows the station layout and calculation order of this case. Values are teaching data.
Case data are teaching assumptions. Stage acceptance follows the equipment manufacturer agreement; a 0.1″ class instrument resolves about 0.02 μm at a 50 mm span, and the achievable uncertainty depends on environment and repeatability.