MEASUREMENT METHOD

Small Angles & Indexing Error

Measure small angular differences or evaluate an indexing table with an appropriate polygon or angular reference.

Principle

An angular reference generates the nominal rotation; the autocollimator measures the remaining small difference.

Preparation

Define the angular range, calibration corrections, measurement axes, target positions and approach directions.

Workflow diagram

01Fix the test item and angular reference and check the 02Set the reference reading and confirm the instrument u03Move through the specified angular positions and recor04Compare readings with reference angles05Report indexing error

Procedure

  1. Fix the test item and angular reference and check the beam return and mounting stability.
  2. Set the reference reading and confirm the instrument units and angular convention.
  3. Move through the specified angular positions and record the direction of approach.
  4. Compare readings with reference angles, applying calibration corrections and repeating the sequence.
  5. Report indexing error, repeatability and reversal effects separately.

Calculation

Error is actual angle minus nominal angle. Apply reference-certificate corrections with signs defined by the measurement model; process both approach directions separately.

Illustrative example

Schematic · not measured data
Schematic: reference, reflecting face and readout, not a performance claim.

Illustration: a corrected position of 10°+0.4″ has a +0.4″ error relative to nominal 10°.

Limitations

  • Resolution is not measurement accuracy.
  • Displays calibrated in mirror angle must not be divided by two again because the beam is reflected.
  • Range, cross-axis sensitivity and reference-surface quality affect results.

Reporting

Retain raw data, datum definition, temperature, repeatability, algorithm and uncertainty. Illustrations are not test results.

Method reference ↗ Method reference ↗ 
STEP BY STEP

Understand the measurement step by step

Principle schematic · not to scale
1 / 4
  1. Fix target angle · Fix the target angle and reference face; define each input as a deviation from nominal rather than a large absolute angle.
  2. Repeat readings · Enter three illustrative deviations with the same settling procedure, retaining every observation.
  3. Apply correction · Apply the defined additive correction c=−0.2 arcsec to each raw deviation and then calculate the mean.
  4. Calculate and check · Calculate a sample standard deviation and identify calibration-correction uncertainty as a separate budget item.

Teaching animation: follow the stations and reference directions. Do not read dimensions or angles from the schematic.

WORKED EXAMPLE

Applying an angular correction to repeated readings

Teaching example with assumed data, not a customer project, instrument accuracy claim or acceptance result.

Example task

Teaching demonstration, not a real customer project: apply a known additive correction to three deviations at one target angle, separating the mean, repeatability and correction sign.

Setup and preparation

  • Fix the target angle and reference face; define each input as a deviation from nominal rather than a large absolute angle.

  • Enter three illustrative deviations with the same settling procedure, retaining every observation.

  • Apply the defined additive correction c=−0.2 arcsec to each raw deviation and then calculate the mean.

  • Calculate a sample standard deviation and identify calibration-correction uncertainty as a separate budget item.

Example data and records

  • Raw deviations: [+0.8,+0.9,+0.7] arcsec.

  • Additive correction c=−0.2 arcsec; corrected values [+0.6,+0.7,+0.5] arcsec.

  • n=3; corrected mean +0.6 arcsec.

Calculation steps

e_corrected=e_raw+c; s=√[Σ(ei−mean)²/(n−1)]=0.1 arcsec. With independent identically distributed readings, the Type A standard uncertainty of the mean is s/√3=0.057735 arcsec, not the total uncertainty.

Interpretation

The worked mean deviation is +0.6 arcsec. A constant correction does not change the sample standard deviation. No complete uncertainty or instrument conformance result is asserted.

Verification and pitfalls

  • Use the correction sign defined by the calibration record; error and correction often have opposite signs.

  • Three observations illustrate the arithmetic; an actual plan must specify the required sample size.

  • Include the reference angle, face quality, environment and correction uncertainty in the full budget.